Operation
ProLab 2000
56
ba75577e03
11/2009
Temperature while
calibrating and
measuring
For precise ISE measurements the temperature difference between
measurement and calibration should not be greater that 2 K. Therefore,
adjust the temperature of the standard and measuring solutions
accordingly. If the temperature difference is greater the
[TempErr]
warning appears in the measured value display.
ISE Cal
This is the conventional
two-point
or
three-point calibration
procedure
that uses 2 or 3 freely selectable standard solutions. The
concentration expected in the measurement determines the
concentration of the calibration standards.
Stability control
During calibration, the stability control is automatically activated.
The current measurement with stability control can be terminated at
any time (accepting the current value).
Calibration record
When finishing a calibration, the new calibration values are displayed
as an informative message (
i
symbol) first. Then you can decide
whether you want to take over these values of the new calibration or
whether you want to continue measuring with the old calibration data.
After accepting the new calibration values the calibration record is
displayed.
Display of calibration
data and download to
interface
You can have the data of the last calibration displayed (see page 87).
Subsequently, you can transmit the displayed calibration data to the
interface, e. g. to a printer or PC, with the
<PRINT>
key.
Note
The calibration record is automatically transmitted to the interface after
calibrating.
Sample record:
Calibration evaluation
After calibrating, the meter automatically evaluates the calibration.
Printing date
26.04.06 16:13
ProLab 2000
Ser. no.
06249876
USER ID
: 1234567890
Calibration ISE
Calibration date
20.04.06 10:14:03
USER ID
: 1234567890
Standard 1
0.010 mg/l
Standard 2
0.020 mg/l
Voltage 1
0.0 mV 24.0 °C
Voltage 2
9.0 mV 24.0 °C
Slope
29.9 mV
Sensor
+++