Operation
ProLab 2000
78
ba75577e03
11/2009
Calibration via a
Comparison meas.
For this calibration procedure, the
Comparison meas.
setting must be
set to
On
in the measuring menu.
Note
Before calibrating via a comparison measurement, the sensor should
be calibrated in the air calibration vessel.
Proceed as follows to calibrate the meter:
8
You have the following options:
z
Accept the new calibration values with
<MENU/OK>
.
Subsequently, the calibration record is displayed and
downloaded to the interface at the same time.
z
To switch to the measured value display without accepting
the new calibration values, press
<MODE>
or
<ESC>
.
i
Relative slope =
0.98
i
++
Accept
LFOx1400-ID <CAL>
1
Connect the D.O. sensor to the meter.
2
Immerse the D.O. sensor in the reference solution.
3
In the measured value display, select the D.O. measuring
window with
<
▲
><
▼
> and
<MODE>
.
4
Start the calibration with
<CAL>
.
The calibration display appears.
i
Immerse sensor in ref. sol.
Continue
LFOx1400-ID <CAL>