background image

Height adjustment column

Operating table OPX mobilis and OPX mobilis RC –

 E

DITION

 

05-03-GB

 ID.-Nr.: 02008857

19

Replacing the check valve and the hydraulic cylinder 
for lateral adjustment

All operating tables can be adjusted by means of 
hydraulic cylinders. The hydraulic cylinders are equipped 
with check valves. Hydraulic cylinders and check valves 
can be replaced.

Preparation

In order to replace one of the check valves or one of 
the hydraulic cylinders, detach the cladding panels of 
the height adjustment column. You will find details 
under “Removing and re-attaching the cladding 
panels” on page 11.

Removing the check valve at the hydraulic cylinder for 
lateral adjustment

Make the hydraulic cylinder retract completely.

Mind the position of the hydraulic hoses at the check 
valve in order not to confound the directions of 
motion.

Loosen the sleeve nuts at the two hydraulic hoses and 
at the two tubes connected to the valve.

Loosen also the sleeve nuts at the tubes at the 
hydraulic cylinder.

Remove the check valve from the hydraulic cylinder.

Installing the check valve at the hydraulic cylinder for 
lateral adjustment

Connect the check valve to the hydraulic tubes and 
retighten all sleeve nuts at the tubes. Mind the tilt 
position of the check valve.

Refasten the hydraulic hoses at the check valve.

De-aerate the system afterwards. You will find details 
in the chapters “Filling new hydraulic cylinders with 
hydraulic fluid and installing them” on page 17 and 
“De-aerating the lateral or Trendelenburg adjustment 
cylinder and the check valves” on page 18.

Check valve for 
lateral adjustment 
cylinder

Hoses

Observe slight 

longitudinal tilt!

Содержание OPX mobilis 200

Страница 1: ...Service manual Operating table OPX mobilis and OPX mobilis RC...

Страница 2: ...manual contains instructions how to carry out repairs on the mobile operating tables OPX mobilis 200 OPX mobilis 300 C OPX mobilis 300 CL OPX mobilis 300 CE OPX mobilis 300 CLE OPX mobilis RC30 OPX m...

Страница 3: ...heck valves and hydraulic cylinder for Trendelenburg adjustment 22 Sealing or replacing the hydraulic cylinder for height adjustment 24 Sensors in the mobilis RC 26 Adjusting the guidings at the heigh...

Страница 4: ...ulic pump 59 Electrical safety 61 Carrying out the measurements 62 Documentation and evaluation of the tests 62 Visual inspection 63 Measuring the resistance of the protective conductor 64 Measuring t...

Страница 5: ...rating tables OPX mobilis also called operating tables in the following Our products are constantly being improved this is why constructional changes carried out after printing of this service manual...

Страница 6: ...nent danger of death or serious injury v Caution This symbol will appear where situations are described which might be dangerous and which might inflict slight injuries a Attention This symbol will ap...

Страница 7: ...extensions or reconstructions of the operating table unless approved by the manufacturer Spare parts have to meet the requirements stipulated by the manufacturer This is always guaranteed when using o...

Страница 8: ...Introduction 8 Service manual...

Страница 9: ...ding on the model All operating table models are mobile They are equipped with four electrically conductive double castors with central locking device The operating tables with a C in their type desig...

Страница 10: ...udinal displacement function By means of this feature such parts of the patient s body can be screened which are otherwise inaccessible to the C arm equipment The operating tables with a G in their ty...

Страница 11: ...ding panel and lift the bellows The four fastening screws at the upper rim of the upper cladding panel are now accessible Detaching the cladding panels The cladding panels consist of two parts which a...

Страница 12: ...anels have to be re attached in reverse order Place the lower cladding panels onto the mobile base and connect them by means of the screws screwed into the connecting strips Fix the cladding panels to...

Страница 13: ...ating table by means of a piece of string Adjust the table top and the back section into horizontal position Lift the cladding of the mobile base up to the lower side of the table top Fix it there usi...

Страница 14: ...ng and installing the control box in mobilis RC table models on page 53 Prepare the necessary tools for jacking the operating table up Removing or lowering the floor pan Loosen the two fastening screw...

Страница 15: ...s out pull the expansion rivets out and lift the plastic cladding off Disconnecting the hydraulic system between the upper section and the lower section for the kidney bridge and for the RC 40 model A...

Страница 16: ...attaching the table top In order to re attach the table top position the table top on top of the table column and adjust it Screw the fastening screws in and tighten them Connecting the hydraulic sys...

Страница 17: ...linder pulling the tube slightly out of its seat The screw joints at the piston rod end are still loose Wait 2 to 5 minutes until the residuary air has ascended inside the piston end of the cylinder P...

Страница 18: ...nd of the cylinder Then loosen the screw joints at the piston rod end of the hydraulic cylinder pulling the tube slightly out of its seat The screw joints at the piston end are still loose Pump hydrau...

Страница 19: ...y Mind the position of the hydraulic hoses at the check valve in order not to confound the directions of motion Loosen the sleeve nuts at the two hydraulic hoses and at the two tubes connected to the...

Страница 20: ...ed by means of circlips Remove the check valve In order to detach the hydraulic cylinder from the upper bearing remove the circlip at the bolt of the upper bearing and start the bolt Lift the table to...

Страница 21: ...raulic fluid has to be topped up as required Top the hydraulic cylinder up with hydraulic fluid You will find details in the chapter Filling new hydraulic cylinders with hydraulic fluid and installing...

Страница 22: ...d the directions of motion Loosen the sleeve nuts at the two hoses and at the two tubes which are connected to the valve In addition loosen the sleeve nuts at the tubes at the cylinder Remove the chec...

Страница 23: ...hydraulic cylinder for Trendelenburg adjustment Install the new hydraulic cylinder You will find details in the chapters Filling new hydraulic cylinders with hydraulic fluid and installing them on pag...

Страница 24: ...nting Detach the four hydraulic hoses of the lateral cylinder and of the Trendelenburg cylinder at their lower ends Loosen the lateral guidings of the height adjustment column You will find details un...

Страница 25: ...ssembly Re assemble the system in reverse order Refasten the lateral guidings with sufficient initial tension as to prevent the column without the table top from sinking by its proper weight You will...

Страница 26: ...k section adjustment inside the seat section frame is covered The operating table is adjusted into horizontal and lowest position by pressing the 0 key on the hand held control unit or on the foot con...

Страница 27: ...the cladding panels on page 11 Re adjusting the lateral guidings On the right hand side of the height adjustment column there are two plastic bearings On the left hand side opposite there are two plas...

Страница 28: ...bearings in such a way that the column glides easily over the whole height adjustment range h It is part of the normal function if individual ball bearings come to a temporary standstill while the inn...

Страница 29: ...ion and for the seat section a Attention When the gas springs have been removed the back section may fall down and thus damage the follower screw of the safety sheet between the seat and back sections...

Страница 30: ...sher at the inside Detach the claddings at the underside of the seat or back section at the back section also the pushing handle h The threaded bolts are secured by means of screw locking varnish They...

Страница 31: ...a hydraulic cylinder The pressure developed is directed via hydraulic hoses to the two trigger heads at the gas springs This way the gas springs are released simultaneously and can be adjusted The tri...

Страница 32: ...gas spring out of the trigger head and pull the gas spring down and out of the back section frame Replacing the hydraulic system The hydraulic system for releasing the gas springs consists of a hydrau...

Страница 33: ...face of the hydraulic fluid Adjusting the level of hydraulic fluid inside the hydraulic system When the hydraulic system has been opened it has to be topped up with hydraulic fluid afterwards There is...

Страница 34: ...tighten the locknut Drive the long bearing bolt from the inside of the back section frame through the trigger head At the same time fix the casings of the hydraulic cylinders of the gas spring release...

Страница 35: ...e mechanism on page 31 Preparation Fold the back section on top of the seat section You will find details in the chapter Folding the back section on top of the seat section on page 29 Removing the hyd...

Страница 36: ...crew joints at the hydraulic cylinder again pulling the hoses slightly out of their bearing Pump the hydraulic fluid in Lowering back section position carefully through the hoses to the hydraulic cyli...

Страница 37: ...ils in the chapter Disconnecting the hydraulic system between the upper section and the lower section for the kidney bridge and for the RC 40 model on page 15 Removing and installing the piloted check...

Страница 38: ...eight adjustment column In its initial state the longitudinal displacement function is locked In this position the teeth of a stop panel are pressed into those of a rack by means of springs The table...

Страница 39: ...e back section on top of the seat section on page 29 Depending on the kind of repair one of the brackets long casting has to be detached You will find details in the chapter Detaching the bracket on p...

Страница 40: ...lders Pull the stop panel together with the bearing elements and the bearing holders off the table top Mind the springs and the tappets which are now loose Installing the stop panel In order to instal...

Страница 41: ...racks are fixed to the underside of the bearing tubes They are accessible from the left or right hand side of the operating table The table top need not be detached Removing the rack In order to remov...

Страница 42: ...The relay arms are now accessible and worn parts can be replaced or loosened connections re tightened Re assembly is carried out afterwards in reverse order Now you can loosen also the two holding scr...

Страница 43: ...udinal displacement function consists of three hydraulic cylinders which are connected by means of a T piece The release mechanism is a closed system For its proper functioning a given quantity of hyd...

Страница 44: ...d repairing the release mechanism on page 31 and following In order to replace the input cylinder proceed as follows Remove the left hand handle together with the cover plate Loosen the threaded bolt...

Страница 45: ...linder into the casing and fix it by means of the circlip or by means of the locking screws at the casing of the actuating cylinder Reassemble the complete system in reverse order h Make sure not to l...

Страница 46: ...y a new strip of adhesive tape Lay in the new sliding strip its bronze coloured back facing downwards onto the adhesive tape Replacing the bearing bushes h The bearing bushes are secured by means of s...

Страница 47: ...ng the old batteries Before replacing the batteries switch the electro hydraulic pump off at the main switch Open the flexible clips securing the supply cables and lift the batteries off their support...

Страница 48: ...nd details under Adjusting the reed board on page 49 Replace the hand held control unit Check the cable connecting the reed board to the hand held control unit Check the cable leading from the control...

Страница 49: ...ection until the LEDs show the correct function Finally secure the reed board by means of the screws and the 8 mm nuts h If the selector lever with the permanent magnet has been lifted off its seat af...

Страница 50: ...ils in the chapter Detaching and re attaching the cladding panels on page 11 Lift the cladding off the mobile base You will find details in the chapter Removing and re attaching the cladding of the mo...

Страница 51: ...the cladding of the mobile base on page 13 Removing the control box The control box is fixed to the side of the floor pan by means of four screws The outer fastening screws are secured to the rear of...

Страница 52: ...box without replacing the cables for the reed board and for the hand held control unit In order to do so loosen the inner fastening screws at the control box and open the box Inside you will find the...

Страница 53: ...ll the control box out of the floor pan Detach the white plug type connector for the motor and the black plug type connector for the batteries Loosen the plugs for the hand held control unit for the m...

Страница 54: ...onal castor can be raised by extending the tie rod In order to detach the clevis loosen the locknut of the clevis detach the circlip of the bolt and pull the bolt out of the clevis and out of the bear...

Страница 55: ...lly in case of a shock Screwing the adjusting screw in will facilitate loosening the brake while screwing the adjusting screw out will make it more difficult to loosen the brake In order to adjust the...

Страница 56: ...of the mobile base on page 13 Remove the floor pan You will find details under Replacing or lowering the floor pan on page 14 Removing the hydraulic switch valve The hydraulic switch valve is fixed to...

Страница 57: ...valve using the hand held control unit first until the air has escaped from the first screw joint Then use the foot pump to pump hydraulic fluid through the switch valve until the air has escaped fro...

Страница 58: ...t of the double castor out Loosen the fastening screw at the bearing lift the table up and pull the double castor down and off v Caution Once one of the castors has been removed the table may tilt and...

Страница 59: ...e pressure hose and to the tank hose by means of banjo bolts In order to remove the electro hydraulic pump switch the main switch OFF detach the electric connections at the lamp wire connector Loosen...

Страница 60: ...Repairs 60 Service manual...

Страница 61: ...ks allow in general the evaluation of the status of the electrical safety After repairs or modifications which may influence the normal operation as to the input output values of the appliance or curr...

Страница 62: ...measurements of the equivalent of the leakage current The appliances may only be operated further if the measured value is lower than the indicated limit value Repeated measurements The values determ...

Страница 63: ...d its connection points After that the appliances are closed if they need not remain open for the tests After the assembly all mechanical parts which have been removed during the repairs must be visua...

Страница 64: ...ch can assume a dangerous electric voltage in the case of a malfunction and the connections of the protective conductor is measured For appliances with mains cable the resistance between the contact p...

Страница 65: ...eakage current is to recognize as soon as possible any deviations from the so called initially measured value and thereby modifications of the safety structure of the appliance These measurements may...

Страница 66: ...e connected with the measuring device according to the following diagram The maximum values for the equivalent of the leakage current amount to 1 mA for appliances with mains cable Measuring the equiv...

Страница 67: ...PX mobilis RC EDITION 05 03 GB ID Nr 02008857 67 Designation of the appliance Type production number Inventory number Testing authority Name Date Kind of testing Results Signature of the inspector Ini...

Страница 68: ...Electrical safety 68 Service manual Designation of the appliance Type production number Inventory number Testing authority Name Date Kind of testing Results Signature of the inspector...

Страница 69: ...Table base Operating table OPX mobilis and OPX mobilis RC EDITION 05 03 GB ID Nr 02008857 69...

Страница 70: ...1 080mm 720mm to 1 080mm 740mm to 1 100mm 720mm to 1 080mm 740mm to 1 100mm 720mm to 1 080mm 740mm to 1 100mm 720mm to 1 080 mm Trendelenburg adjustment 30 Reverse Trendelenburg adjustment 30 Lateral...

Страница 71: ...20mm F1 10A slow blow type F2 2A medium time lag type Battery type employed 4 pcs 6 V 13 5 Ah Input AC of the battery charger 117V or 230V 50 to 60Hz 0 4A Output DC of the battery charger 27 6V 1 4A...

Страница 72: ...ronics cable Reed board Control unit Hand held control unit External fuse External fuse Charging socket Set of cables and batteries 708058 Cabling of the motor and of the battery illustrated position...

Страница 73: ...Lowering valve with integrated throttle 02011260 Double piloted check valve 02009507 Lateral adjustment cylinder 02009508 Trendelenburg adjustment cylinder 02009509 Height adjustment cylinder 02011223...

Страница 74: ...cal data 74 Service manual Overview electric system mobilis 300 E Overview electric system mobilis 300 E mobilis control unit Reed board Hand held control unit green yellow EMC protector motor EMC pro...

Страница 75: ...Overview electric system mobilis RC30 40 Overview electric system mobilis RC 30 40 Control box mobilis RC30 40 Battery Cable control box connector socket Hand held control unit Connector box sensors C...

Страница 76: ...down Bypass foot pump Contacts per connection Terminal box Terminal nos Cabling of the cable terminal box D Sub 25 pole Core colour Terminal no Pin 1 white 2 Pin 2 brown 1 Pin 3 green 3 Pin 4 yellow 4...

Страница 77: ...ight adj up Height adj down Bypass foot pump Contacts per connection Terminal box Terminal nos Cabling of the cable terminal box D Sub 25 pole Core colour Terminal no Pin 1 white 2 Pin 2 brown 1 Pin 3...

Страница 78: ...bling of the sensors mobilis RC30 40 Cabling of the sensors mobilis RC30 40 Contacts per terminal D Sub 25 pole plug Assignation of the terminals 1 Height adj 2 Trendelenburg adj 3 Lateral adj 4 Back...

Страница 79: ...lve compl 02011271 Slide valve compl for 200 type models 02011266 Lowering valve 02011260 Double piloted check valve 02011220 Double piloted check valve for E type models 02009507 Lateral adjustment c...

Страница 80: ...63 Magnetic valves 02010805 Screw joint with filter 02013112 Hydraulic hose 02013116 Hydraulic hose 02011260 Check valve lateral adj 02013020 Check valve Trendelenburg adj 02012332 Lowering valve 0200...

Страница 81: ...alve 02012351 Magnetic valves 02010805 Screw joint with filter 02013111 Hydraulic hose 02013112 Hydraulic hose 02011260 Check valve lateral adj 02013113 Hydraulic hose 02013116 Hydraulic hose 02013220...

Страница 82: ...n up down up Kidney bridge Back section adj RC40 only Tables with longitudinal displacement function Tables without longitudinal displacement function 2015102 02012482 Foot pump compl 02014894 Switch...

Страница 83: ...diagrams exploded views repair instructions or spare parts lists can be obtained upon request from Schmitz u S hne When ordering technical descriptions or spare parts at the manufacturer s kindly ind...

Страница 84: ...4162 http www schmitz soehne de e mail export schmitz soehne de Branch offices Spain SCHMITZ u S hne Ib rica S L Avda de Bruselas 58 semis tano E 28028 Madrid Phone Fax 34 91 7130000 Switzerland SCHMI...