3.5 Error Messages
Error message 1
:
- The DISPLAY shows
EEE
.
The upper limit of the tension range was exceeded by more than 10%.
Reduce the line tension.
OR
AUTO ZERO is no longer possible.
Recalibrate the instrument following the directions in Chapter 3.6.1.
Error message 2
:
- The DISPLAY shows
.
The lower limit of the tension range was fallen below by more than 10%.
Properly insert the process material.
OR
AUTO ZERO is no longer possible.
Recalibrate the instrument following the directions in Chapter 3.6.1.
3.6 Static Verification of Measuring Accuracy
All tension meters are calibrated with standard materials - such as polyamide monofila-
ment (PA) - according to the SCHMIDT factory procedure. The diameters are given in
Chapter 2.Any difference in process material size and rigidity from the standard material
may cause a deviation of the accuracy.In 95% of all industrial applications, the SCHMIDT
calibration has been proven to provide the best results and is used for comparative pur-
poses.The tension meter is factory calibrated for a vertical material path (fig. 3.6.a).