Technical Data
12
User guide
ECG
12.2
CARDIOVIT AT-1 G2
12.2 ECG
Environmental conditions EMC
• IEC/EN 60601-1-2
• CISPR 11/32 Group 1, class B
The
CARDIOVIT AT-1 G2
can be exposed to the following interferences without any
impairment of the essential performance features:
• static discharge up to ±8 kV contact, ±15 kV air
• radio frequency range up to 10 V/m (80 to 3000 MHz)
• near fields of wireless HF telecommunication devices 9 to 28 V/m (385-5785
MHz), for more details, see table on page
54
.
• Magnetic fields of 30 A/m, 50 Hz
Patient input
Fully floating and isolated, defibrillation-protected (only with original SCHILLER pa-
tient cable)
Lead configurations
• Standard 12 channels
Display
Leads
Status
• Display of selected leads (3 x4)
• Filter status
• Power source
• Leads
• Electrode contact status
• Heart rate (HR)
• Date and time
• Patient number
Filter
Myogram filter (muscle tremor)
Notch filter
Set to 25, 40, 150, 250 Hz (250 Hz = Filter Off)
Distortion-free suppression of superimposed AC 50 or AC 60 Hz sinusoidal interfer-
ences by means of adaptive digital filtering
Data record
With optional interpretation ETM
• Listing of all ECG recording data (date, time, filter)
• ECG measurements results (intervals, amplitudes, electrical axes)
• Averaged complexes
• Guidance on interpreting adult and paediatric ECGs
ECG amplifier
Complies with IEC 60601-2-25 and ANSI/AAMI EC11