6
Resting ECG
6.4
Changing the ECG display
Page 34
CARDIOVIT AT-1 G2
A
rt.
no.
: 2.
511
235 Rev.:
d
6.4 Changing the ECG display
6.4.1
Display
Leads
The following presentation can be selected in
Menu > Settings > ECG > Leads
& cable
:
The Standard and Cabrera lead sequences are as follows:
6.4.2
Myogram filter
The myogram filter suppresses disturbances caused by strong muscle tremor. In
Menu > Settings > ECG > Filters &
formulas,
the
myogram filter is
defined.
6.4.3
Other filters
The following additional filters are available:
Baseline filter
The cut-off frequency for the baseline filter is based on IEC 60601-2-25 and cannot
be changed.
Notch filter
This filter prevents recording interference due to mains frequency oscillation. If the
filter is active, "AC 50 Hz" or "AC 60 Hz" is displayed.
The ECG display is optimised for one column of 3 channels and cannot be changed.
The amplitude and speed can be changed at any time with the direct keys. The
standard values for the display and thermal printer
are
25 mm/s
and
10 mm/mV
.
Lead sequence
Lead group 1
Lead group 2
Lead group 3
Lead group 4
Standard
I, II, III
aVL, aVF, -aVR
V1, V2, V3,
V4, V5, V6,
Cabrera
aVL, I, -aVR
II, aVF, III
V1, V2, V3,
V4, V5, V6
In the information field, LP 25 Hz, LP 40 Hz ,
LP 150
Hz or OFF is displayed.
The cut-off frequency is user-defined at LP 25
Hz, LP 40
Hz or
LP 150
Hz or LP Off
(250
Hz) (
see chapter 9.2.2
, page
41
).
25 mm/s 10 mm/mV LP 150Hz AC 50 Hz
• The filters are activated/deactivated or changed in the ECG settings (
see following
description
).