Page 37
A
rt.
no.
: 2.
511
235 Rev.:
d
Culprit Coronary Artery Algorithm
7
User guide
Introduction
7.1
CARDIOVIT AT-1 G2
7.1.2
Starting the CCAA analysis
Procedure
The data is shown in the print preview. The recording can be checked, accepted and
further printouts obtained in different formats, and it can be exported as PDF.
When the CCAA analysis option is activated, make sure that the C4 electrode is
attached in position C4r (precordial), see
section 4.4 Standard with C4r for CCAA
recordings, page 24.
1. Enter the patient data.
2. Activate
CCAA
by selecting “Yes”.
3. Enter the additional parameters Bypass/stenting and time since chest pain
started.
4. Check the electrode placement (V4r) and record the ECG.
Patient ID
. .
Gender
Pacemaker
Patient data
25.01.2018 17.58
Female
No
Date of Birth
next/previous
select
close
Activate CCAA
Yes
Prior Bypass/Stenting
No
Chest pain for [h]
0.0
All other settings and features (saving, printing etc.) are the same as described in
section 6.2 Automatic resting ECG recording, page 31
.