3
Operation
3.7
Procedure in case of an alarm
Page 22
A
rt.
no.
: 2.
510
474 rev.
: d
ARGUS LCM/PLUS
3.7 Procedure in case of an alarm
3.7.1
Display of alarms
During initial switching on
The alarms are suppressed for a defined time (programmable in the menu
About+/
Alarm.
Alarm suppr. time standard is 3 minutes)
The message
Alarm suppressed 3:00
appears.
During monitoring
There are two alarms:
• Technical alarm, displayed in the alarm status field on the top right.
In the case of a technical alarm, an audible alarm sounds and the measurement
field for the respective value flashes.
• Physiological alarms, displayed in the alarm status field on the top left.
In the case of a physiological alarm, an acoustic alarm sounds and the measure-
ment field for the respective value flashes.
3.7.2
Switching off an alarm
Each alarm off function is password protected. Read carefully the following warning
and information:
Entering alarm OFF password
V
When the acoustic alarm is switched off, the patient acoustic physiological
alarms are silenced and suppressed indefinitely. Use this function only if discon-
necting a sensor from the patient for a long period of time.
• When the
alarm off
button is pressed, the audible alarm for the respective param-
eter is suppressed. This is indicated by the LED above the button and a message
on the monitor, e.g.
Alarm ECG OFF
. If an alarm occurs as long as the
alarm off
button is pressed, a visual alarm is displayed in the respective measurement field.
• With the LCM plus version E the access to the NIBP and IBP alarm off function will
be executed by the NIBP alarm button. A menu with blood pressure alarm off (com-
mon for NIBP and IBP) and IBP calibration function appears.
BP Alarm On/Off
Connect zero mmHg
Start IBP calibration
End / Escape
1. Press the desired alarm off button. The password dialog appears.
2. Press following button to enter the password:
,
, ,
1. The password protection can be disabled in the special menu see chapter
page
Enter Password
Press [Enter] to cancel
I,II, III