7
Appendix
7.1
Settings menu
Page 44
A
rt.
no.
: 2.
510
474 rev.
: d
ARGUS LCM/PLUS
7.1.2
Special menu “About+”
V
Only authorised personnel trained in the operation of this device are allowed to
do the setups in this menu.
The extended menu
About+
is displayed by pressing the following button combina-
tion:
Press ;
select
About
; press
and then
Main menu extended
Parameter
Value
About+
Alarm
Alarm stop mode
On/
Off
Alarm suppr. time
2
min (2 ... 10)
Alarm sound
DIN EN 475
/Standard
Nurse call
Immediately/
in 5 sec.
Alarm restriction (see page
On/
Off
Param
Lead
II
Lead speed
12.5/
25
/50 mm/s
Respiration source
a
ECG
/CO
2
a. When the set respiration source is ECG and the pacemaker is on, respiration measurement via the ECG is not possible. The respi-
ration wave field is not displayed in this case.
Respiration speed
6.25
/12.5 mm/s
SpO
2
Pulse
4/
8
or 16 s
SpO
2
sensitivity
Low
/High
NIBP adult Initial
180
NIBP pediatric Initial.
150
NIBP neonatal Initial
120
NIBP MAP
On
Temperature unit
°C
/°F
CO2
mmHg
/kPA/Vol%
Trend
View time
1, 5, 15,
30,
60, 120, 240 min
Device
Language
English
/Deutsch/
b
Français/Svenska/American/Ital-
iano/Español/Portuges/Norge
b. The Russian language requires a separate software.
Software
Save as default
With this function, changed values can be perma-
nently saved. (See section
Restore defaults
With this function, the saved default values can be re-
stored. (See section
Factory defaults
Restores the SCHILLER factory settings. The display
language is changed to German. (See section
.
Service
For service only
NIBP update or ECG electrode test.
Program update
For service only