7.3 Connecting the ECG patient cable
7.3.1
3-leads cable for children and neonates
7.3.2
3-lead cable for respiration
7.3.3
3-lead cable for adults
Important
The guidelines for patient electrode placement are provided as an overview only.
They are not a substitute for medical expertise.
RRi
An RRi monitoring with HF-ECG cable against high frequncy signals is not possible.
The used HF-signal for the RRi measuring will be filtered by the HF protection in the
ECG cable.
Pacemaker monitoring with HF-ECG cable
Pacemaker monitoring with HF-ECG cable against high frequncy signals is not pos-
sible. The used pacemaker impulse will be filtered by the HF protection in the ECG
cable.
V
Danger of destroying the device during defibrillation! The device is only type CF
protected if the original SCHILLER patient cables are used.