Instructions for Use

multisonic

®

Inhaling with ultrasonic

top seiten engl.  03.04.2003  15:52 Uhr  Seite 1

Содержание Multisonic MN 80170

Страница 1: ...Instructions for Use multisonic Inhaling with ultrasonic top seiten engl 03 04 2003 15 52 Uhr Seite 1 ...

Страница 2: ...for socket Socket 12V DC Baffle plate Rechargeable battery pack MN 81300 depending on design Nebuliser head Inhalation valve dark blue Function display green yellow charging status display yellow Button for capacity test Plug 12V DC Plug in mains supply unit MN 81200 Mouthpiece top seiten engl 03 04 2003 15 52 Uhr Seite 2 ...

Страница 3: ...tion and sterilization 13 Use of the rechargeable battery pack 19 Trouble shooting 21 Information on medicines 22 Spare parts consumable 23 Optional accessories 24 Service information and guarantee conditions 25 Explanation of symbols 26 Technical data and classification 27 Table of Contents top seiten engl 03 04 2003 15 52 Uhr Seite 3 ...

Страница 4: ...se of medi cines into the environment The comfort of patients was one of the main con siderations in the design of the multisonic top Universal usage Optimal penetration of tissues High nebuliser performance and short inhalation times Warming of aerosol Low wastage of medicines and ease of setting doses Simple and safe to use Easy to clean Noiseless in operation Modern design Small easy to handle ...

Страница 5: ...wing components 1 hand unit 1 nebuliser head 1 mouthpiece 1 baffle plate 1 inhalation valve 1 metering aid 1 plug in power supply unit 1 carrying case 1 set of Instructions for Use Guarantee form top seiten engl 03 04 2003 15 52 Uhr Seite 5 ...

Страница 6: ...ug must be pulled out of the socket on the rear side of the unit and re inserted to re start the nebu liser The multisonic top can be used for nebulisation of all Inhalation solutions that are water based including suspensions and natural therapy agents that are suitable for inhalation Use only medicines that have been prescribed or recommended by your physician If you have inhaled the wrong medic...

Страница 7: ...iple of function Air inflow Inhalation valve Baffle plate Medicine Ultrasound waves Ultrasonic generator quartz Aerosol exit Air and particles Aerosol ON OFF Button top seiten engl 03 04 2003 15 52 Uhr Seite 7 ...

Страница 8: ...erage pulmonary deposition NaCl 0 9 20 15 10 5 0 0 2 4 6 8 Duration of inhalation secs Inhalation Breathing in slowly deeply Pause in breathing Hold breath Exhalation Exhale deeply Yellow LED is lit top seiten engl 03 04 2003 15 52 Uhr Seite 8 ...

Страница 9: ...socket Remove the nebuliser head from the hand unit and place the hand unit on a flat surfa ce such as a table Ensure that all parts of the unit are fully dry before commencing the inhalation process Using the multisonic top for the first time Please note The multisonic top must not be used in rooms such as bathrooms that have a high humidity level because of the danger of electric shock 230 VA to...

Страница 10: ...rrectly in the groo ve as shown in Fig 2 Place the nebulizer head on the hand unit Fig 3 by placing Inhalation resting it on the inner edge of the side of the hand unit indicated by the arrow tilting it backwards and snapping it into place Always hold the hand unit vertically Ensure that the medicine chamber always has medicine solution in it Some drops of the medicine will cling to the walls of t...

Страница 11: ...w that the unit is operational Fig 5 A few seconds after the unit has been switched on the aero sol mist forms below the baffle plate The mist remains within the unit until you inhale Bring the hand unit up to your mouth Insert the mouthpiece between the front teeth and put your lips around it Upon inhalation the correspon ding quantity of aerosol will directly penetrate the airways Inhale the aer...

Страница 12: ...e the medicine has been used up down to the residual quantity or because the over heating protection has been triggered then the plug has to be pulled out of the socket on the rear of the unit and re inserted before the unit can be used again Examples of various uses of the multisonic top Fig 1 Fig 2 Inhalation via the mouthpiece Inhalation via a mask Caution Children and persons who need care sho...

Страница 13: ...ffle plate and valve may be boiled or autoclaved at up to 134 C It is recommended that users in Germany use one of the disin fectants listed in the current DGHM list Preparations based on compounds that release halogens strong organic acids and oxygen releasing com pounds are not suitable as they may damage the material of the inhalation unit German Society for Hygiene and Microbiology Please note...

Страница 14: ... on the grate of the cen tral section Now place the central section of the vaporizer on the base section Continued on the next page Disinfection in a vaporizer Caution Make sure that the socket of the 12 V inlet of the multisonic top is sealed with the stopper Before using the vaporizer for the first time it is very important that you read the Instructions for Use Fig 1 top seiten engl 03 04 2003 ...

Страница 15: ...e stopper the hand unit can be disinfected sterilized through immersion disinfec tion placing it in a chemical thermal disinfector 60 C for 10 minutes or in a gas sterilizer The nebuliser head valve mouthpiece and baffle plate can be disinfected gas steril ized or autoclaved at up to 134 C as described above After cleaning and disinfection of the unit in a vaporizer dishwasher or after chemical th...

Страница 16: ...Chemical thermal 60 ºC for 10 mins dishwasher Steam in vaporizer Auto claving at 134 ºC Form aldehyde Plasma Unit X X X X X X X X X X X X X X X X X X X X X X X X X X X X X X X X X X X X X X X X X X X X Mouthpiece Nebuliser head Baffle plate Valve X X X Mask x this option may be used Cleaning Disinfection Sterilization top seiten engl 03 04 2003 15 52 Uhr Seite 16 ...

Страница 17: ...tery pack will light up once it is con nected to the power supply Once charging commences the yellow LED 2 will flash several times The yellow lamp will go out once the battery pack is fully charged Fig 1 After around 3 seconds the LED will light up continuously through out the entire charging period If the battery pack has only a very low charge then the charg ing cycle will commence with a pulse...

Страница 18: ...ectly the bat tery pack can be charged at least 500 times As with all Ni MH batteries the battery pack will only deliver its full power if it has been fully charged If the battery pack is not char ged correctly then its capacity will be reduced It should not be stored above 40 ºC The battery pack should be stored in a cool dry place when not in use Note If not used a fully charged battery pack wil...

Страница 19: ... ensu re that the nebuliser performance is optimal Has the medicine chamber been treated with grease or a detergent in error If so rinse it out thoroughly with water that is warm hot approx 40 C Has an air bubble developed in the ultrasonic generator If so remove it carefully using a blunt instrument Has the overheating protection been triggered Check the quantity of medicine Allow the unit to coo...

Страница 20: ...ent Bero tec Berudual Brom hexin Inhalat Cromoglicin ratioph arm DNCG Stada Emser Sole Eucasil Mibrox Mucosolvan Sultanol sur factant amilorid Pulmicort Colistin Ilo medin Tobra mycin Pulmozym Dosage The unit is not designed for emergency life saving meas ures Respiratory disorders that can be treated with the multisonic top include the following bronchial asthma chronic obstructive bronchitis bro...

Страница 21: ...nfection They must be changed at the latest after they have become unsightly or exhibit dama ge The parts should be replaced as follows Every 3 months Filter holder Nebuliser head Baffle plate Mouthpiece Inhalation valve Please use the attached order card for re ordering consumables Nebuliser head 70574 Mouthpiece set 70630 6x Baffle plate set 85750 3x Inhalation valve 85723 top seiten engl 03 04 ...

Страница 22: ...24 Accessories Mask 70620 top seiten engl 03 04 2003 15 52 Uhr Seite 22 ...

Страница 23: ... 24 months from the date of purchase The multisonic top inhalation unit is an electrical ly operated unit It must not be used without super vision in potentially explosive atmospheres or in rooms with a high humidity Those using this unit must have an exact knowledge of these Instructions for Use and must adhere to the requirements presented in them The unit must only be repaired by the manufactur...

Страница 24: ...er Y Caution observe documentation Type 1 usage component Central positive connection DC 12 V Mark of conformity protected by law Do not use in humid environments Not to be worked on by unauthorized personnel Symbol DC AC IP XY Explanation of symbols 0366 top seiten engl 03 04 2003 15 52 Uhr Seite 24 ...

Страница 25: ...µm Plug in power supply unit MN 81200 Input voltage AC 230 V Output voltage DC 12 V Output current 1 0 A Rechargeable battery pack MN 81300 Input voltage DC 12 V Output voltage DC 12 V Ah capacity 1 8 Weight 0 5 kg Classification Intermittant operation Class of protection II upon operation with plug in power supply unit MN 81200 Disinfection sterilization as described on pages 14 15 MMD class IIa ...

Страница 26: ...wiesen 12 D 07330 Probstzella Phone 49 0 3 67 35 4 63 0 Fax 49 0 3 67 35 4 63 44 info multisonic de www multisonic de Kommunikationsbüro Ulm 0403 We reserve the right to make technical alterations top seiten engl 03 04 2003 15 52 Uhr Seite 26 ...