10 Care and control tasks for operators
D21518 Edition 2017/07
Original Instructions
26/32
10 Care and control tasks for operators
General information
The
motorhandpiece is designed for maximum durability. Despite this, the
chuck should be removed and cleaned from time to time, and the dirt underneath the
handpiece tip removed with a brush.
Contamination on the plastic surfaces can be removed with a dry cloth.
Do not use aggressive cleaning agents!
Care and inspection plan
Care and inspection plan for QUBE II handpiece:
Interval
Care and inspection tasks
Remarks
Weekly
Dismantle the tip and remove dirt
Do not use compressed air!
Once a
month
Dismantle the chuck, clean and lubricate
Lubricate with Art.No. 51/1
E0804/ E0201
No handpiece connected
Check whether the handpiece is
correctly connected - check plug
with bayonet connector!
E0806
Overload cut-off function
Decrease contact pressure on
workpiece.
E0802
Handpiece blocked
Check whether the chuck is closed.
Check smooth running of the
shaft.
Broken cable
Replace the handpiece cable.
General error
The control unit is flashing red,
no display on the control satel-
lite, keys without function
Check connection cable from the
control satellite, remove the mains
plug and plug it back in again.
General error
Keys are not very responsive
Unplug the mains plug and plug it
back in again.
Do not touch the control satellite
when restarting, so as to avoid
miscalibration!
Содержание QUBE II
Страница 1: ...Ausgabe 2017 04 Betriebsanleitung Schick GmbH Original ...
Страница 32: ...12 Anhang D21518 Ausgabe 2017 04 QUBE II Betriebsanleitung 32 34 ...
Страница 35: ...D21518 Edition 2017 07 Original Instructions Original Instructions Schick GmbH ...
Страница 67: ...Édition 2017 04 Notice originale Schick GmbH ...
Страница 99: ...Versione 2017 04 Originale Istruzioni per l uso Schick GmbH ...