CDR Wireless Installation
2.3. Checking for RF Noise
1. At the Wireless Utility, click Start.
2. Click RF environment.
3. Click
Live.
4. Click
Scan.
5. After approximately 2 – 3 minutes, click Stop Logging.
6. Review the results and write down the recommended RF channel(s).
7. Close the RF environment window.
2.4. Performing Final Checks
1. Click the Sensor / Receiver Read info buttons. Verify that the currently set RF
channel is recommended. If not, click the Sensor arrow to select a recommended
channel and click the Match Channels checkbox to automatically set the Receiver.
Both the Sensor and Receiver must be set to the same RF channel to communicate
with each other.
2. Remove Sensor from cradle, attach battery pack, and turn Sensor on. If you need
more information on how to do this, refer to
Section 4
3. Place Sensor on patient chair.
4. Click Sensor monitor and check battery life, signal strength, and RF noise. For more
information about the Sensor Monitor, refer to
Section 5
5. Place Receiver where signal strength is highest and RF noise is lowest. Refer to Base
Station positioning guidelines and options in
Section 6
2.5. Loading the Wireless Sensor Calibration File
For this procedure you will need the CDR Wireless Calibration File disk provided with
your system. This disk contains a unique calibration file for your Wireless Sensor.
1. Start CDR Dicom software.
2. Insert Sensor calibration diskette.
3. Click System > Options and select the X-ray acquisition module.
4. Click Properties and select the Calibration tab.
5. Click Install New Calibration File.
6. Select the calibration file to install and click Open.
7. When the file has been copied to your system, click OK.