Installation and Setup
2.1 Mechanical Installation: Mounting Options and Dimensions
Casing type AT1
Mounting Options
• The best fixing point of the camera is the collar
for the mounting bracket SK5105 (available as an
accessory).
• Four threaded holes M3 x 6.5 mm provide further
options for customized brackets.
• The length and weight of the optics might be beyond
the capability of the standard mounting bracket
SK5105. For this purpose, a second mounting
bracket type SK5105-2 to hold the tube extension
ring(s) is more appropriate.
Optics Handling
• If the camera and the optics are ordered as a kit,
the components are pre-assembled and shipped
as one unit. Keep the protective cap on the lens
until the mechanical installation is finished.
• If you must expose the sensor or lens surface,
ensure the environment is as dust-free as possible.
• Gently blow off loose particles using clean
compressed air.
• The sensor and lens surfaces can be cleaned with
a soft tissue moistened with water or a water-based
glass cleaner.
Mounting bracket SK5105
Mounting system SK5105-2
for cameras with a tube
extension > 52 mm
Clamping set SK5102
Set of 4 pcs. clamping claws
incl. screws
66
50
20
16.5
3.5
6.5
10 10
36
M3
Ø3.3
Ø4.3
6
15
50
.3
41
.7
70 63 40
Ø 42
M4
1/
4’’
20G
6
36
Ø 42
70
3.5
31.5
25
10
3.5
70
63
40
1/
4’
’ 20G
M4
Ø4.3
Hex socket head screw
DIN 912–M3x12
Clamping claw
Pixel 1
M3 (4x)
depth 6.5 mm
CCD-Sensor
52
Ø65
C-Mount
2.5
6
FFL
41.7
Ø42
Lens mount:
C-Mount
Seat for bracket:
Ø42 mm
Flange focal length:
FFL = 17.5 mm
11.1
AT1