SCANMAX 25 C
TROUBLE SHOOTING
1 Insert key and turn on
power.
Is Power light lit
No
Check that the mains lead is connected.
At the wall socket and then to the rear
connection panel.
Check that the mains socket is active.
If not contact electrician
Check mains input fuse
If faulty, replace,
If fuse blows again check with door open
If faulty, replace,
fuse blows again check with door closed
Check bulb
If faulty, replace,
Change controller and x-ray generator
If faulty, replace
2
System Ready
Is System Ready light
on
No
Check door is fully closed
Check Rear panel interlock
Check top panel interlock
Check LED
)
)
) Ensure interlocks are made
)
)
2 Press X-ray On switch
No Picture
Is X-ray On light lit No
Is System Ready lamp lit
Check door is fully closed.
)
) Ensure interlock is made.
)
Check LED
If faulty, replace.
Check mains input fuse
If faulty, replace.