TE-BA-e-2020
9
In general, the larger the LN value, the smaller the reading result for the same layer
thickness. A small change in the LN value causes a large change in the reading result
in the upper measuring range (at 500µm/ 20mil).
The LN value must be corrected in this way:
Pressing the on/off button: It takes about 11 seconds from the start of pressing this
button.
Its value can be changed by pressing the Plus or Minus key after 'LN' appears in the
display and releasing the On/Off key. The value is stored and then the zero key is
pressed.
A. The reading result in the lower area is corrected by pressing the Plus or Minus key.
E.g. the LN value is increased if the reading result is correct in the lower range (e.g.
51µm) but too large in the upper range (e.g. 432µm).
In contrast, the LN value is reduced if the reading result is correct in the lower range
(e.g. 51µm) but too small in the upper range (e.g. 432µm).
C. This procedure of A. and B. is repeated until the reading result is satisfactory in its
accuracy for each distance foil.
Note: