6
TE-BA-e-2020
5.4 To take the next measurement, simply add more
is raised by 1cm, the display shows "0" again and step 4.3 is repeated.
5.5 In case of possible inaccuracies in the measurement result, it is recommended to
adjust the measuring instrument before the measurement as described in chapter 6.
5.6 The device can be switched off with the on/off button.
On the other hand, it switches itself off 50 seconds after the last keystroke.
5.7 The measuring unit can be displayed in µm or mil:
- By pressing the changeover key 4-7
or
- By pressing and holding down the on/off button 4-6 until "UNIT" appears in the display
and then pressing the zero button 4-3.
This process takes a total of 7 seconds (from pressing the on/off button).
5.8 To switch to the measuring mode from "simple" to "continuous" or vice versa, either
- press the SC- button 4-10
or
- the on/off button 4-6 is pressed and held until SC appears on the display. Then the
zero key 4-3 is pressed. The symbol
((-))
stands for the continuous measuring mode
and 'S' for the simple measuring mode.
This process takes 9 seconds (from the time the on/off button is pressed).
6.
Adjustment
6.1 Zero adjustment: The zero adjustment for "F" and "N" should be carried out
separately.
The iron zero plate should be used when the display shows "F" and the aluminium zero
plate when "N" appears in the display.
The measuring probe 4-1 is now carefully placed on the zero plate. The zero key 4-3
is pressed and "0" is shown in the display without lifting the probe.
Attention: The zero adjustment is useless if the measuring probe is not directly
on the zero plate or another uncoated standard material.
6.2 A suitable adjustment foil must be selected according to the typical measuring
range.
6.3 The selected adjustment foil is placed on the zero plate or the uncoated standard
material.
6.4 Carefully press the measuring probe onto the spacer foil and lift it off again. The
reading result appears on the display. This can be corrected by pressing the plus key
4-4 or the minus key 4-5. To do this, however, the measuring probe must be removed
from the zero plate or the measuring body.