TC-BA-e-2020 

 

a)  When  the  battery  symbol  ``+/-``  appears  on  the  display,  the  batteries  should  be 
replaced. 
b) The battery cover 4-8 is removed from the meter and the batteries are removed. 
c)  The  batteries  (4x1.5  V  AAA/UM-4)  are  inserted  by  observing  the  polarity  when 
inserting them. 
d) If the device is not used for a longer period of time, the batteries should be removed. 

8. 

Adjusting foils 

 
This  instrument  includes  an  adjustment  foil  set  with  different  foils  and  measuring 
ranges, but always covers the measuring range from 20 to 2000µm. These are also 
available as optional accessories, article ATB-US07 

9. 

General notes 

 
a) The measuring instrument should always be adjusted on the base material used for 
the actual measurement, if possible, instead of on the supplied zero plate F made of 
steel or N made of aluminium. This makes the measuring accuracy more accurate from 
the outset. 
b) The measuring probe may wear out. The lifetime of it usually depends on the number 
of measurements and the roughness of the layer to be measured. The replacement of 
a measuring probe should only be carried out by qualified personnel  
can be carried out. 

10. 

Restore factory settings 

 
a) It is recommended that the factory settings be performed in the following cases: 
- no more measurements can be made at all 
-  the  measuring  accuracy  is  impaired  due  to  drastic  changes  in  the  environment  in 
which the measuring instrument is used 
- After replacing the measuring probe 
(b) Procedure: 
The factory settings include both the settings for "Fe" (F) and for "NFe" (N). 
Please note which symbol is shown on the display. 
Only one setting at a time can be made, as follows: 
-  Press  the  on/off  button  4-6  and  do  not  release  it  until  the  display  shows  "CAL".  It 
takes about 5 seconds from the start of this button operation until CAL is displayed.  
The measuring probe 4-1 is carefully pressed onto the corresponding carrier material. 
Then the zero key is pressed, which is accompanied by a beep. 
When NF:H appears on the display, the measuring probe 4-1 is raised by more than 5 
cm. 
Then the zero button is pressed again and the meter returns to the measuring mode. 
The factory settings are now complete. 
Note:  This  process  should  always  be  completed  within  the  following  6  seconds. 
Otherwise, it will be automatically interrupted by the unit and the reset will be invalid. 

Содержание TC 1250-0.1F

Страница 1: ...2336 Balingen e mail info kern sohn com Phone 49 0 7433 9933 0 Fax 49 0 7433 9933 149 Internet www sauter eu Instruction Manual Digital Coating Thickness Gauge SAUTER TC Version 2 0 04 2020 GB TC BA e 2020 PROFESSIONAL MEASURING ...

Страница 2: ...lity of this measuring device and its extensive functionality For any questions wishes and suggestions please do not hesitate to contact us Table of contents 1 Introduction 3 1 1 General description 3 2 Functions 3 3 Technical data 3 4 Description of the control panel 5 5 Measuring procedure 5 6 Zero adjustment 6 7 Battery replacement 6 8 Adjusting foils 7 9 General notes 7 10 Restore factory sett...

Страница 3: ... etc This probe is often used to measure the thickness of galvanized layers lacquer layers enamelled porcelain layers phosphorescent layers copper plates aluminium plates alloys paper etc The N probe measures the thickness of non magnetic layers on non magnetic metals It is used to measure anodizations lacquer layers glazes colours enamel plastic layers powder coating etc These should be on non ma...

Страница 4: ...ries UM 4 Ambient temperature 0 C to 50 C Air humidity 80 Dimensions 126 x 65 x 27mm Weight Ca 81g without batteries Scope of delivery Carrying case operating instructions Depending on the type 1 built in measuring probe F TC 1250 0 1 F with F N key 2 built in measuring probe N TC 1250 0 1 N with F N key 3 built in measuring probe FN TC 1250 0 1 FN without F N u S C key Spacer foils for each model...

Страница 5: ...tself by the symbol Fe F for ferrous metals or NFe N for non ferrous metals shown on the display It switches to the automatic measuring mode which also correctly recognizes and assigns the zero plate or any other substrate itself b Place the measuring probe 4 1 on the layer to be measured The layer thickness can now be read on the display The reading result can be corrected by pressing the plus ke...

Страница 6: ...ressing the on off button 6 Zero adjustment a The zero adjustment for Fe F and NFe N should be done separately The iron zero plate is taken when Fe is read on the display while the aluminium zero plate is taken when NFe is read The measuring probe 4 1 is now carefully placed on the zero plate The zero on off key 4 6 is pressed and 0 is shown in the display without lifting the probe Attention The z...

Страница 7: ...probe should only be carried out by qualified personnel can be carried out 10 Restore factory settings a It is recommended that the factory settings be performed in the following cases no more measurements can be made at all the measuring accuracy is impaired due to drastic changes in the environment in which the measuring instrument is used After replacing the measuring probe b Procedure The fact...

Страница 8: ...ng the on off button It takes about 11 seconds from the beginning of pressing the on off button Its value can be changed by pressing the Plus or Minus key after LN appears in the display and releasing the On Off key The value is stored and then the zero key is pressed A The reading result in the lower area is corrected by pressing the Plus or Minus key B The Ln value is increased if the reading re...

Отзывы: