TC-BA-e-2020 

        
d) In case of possible inaccuracies in the measurement result, it is recommended to 

adjust the measuring instrument before measurement, as described in chapter 6. 

 
e)  On  the  one  hand,  the  device  can  be  switched  off  with  the  on/off  button  4-6. 

Otherwise, it switches itself off 80 seconds after the last key is pressed. 

 
f)  The  measuring  unit  can  be  displayed  in  µm  or  mil.  The  changeover  is  made  as 

follows: 

    - By pressing the changeover key 4-7 

or

 

    - the on/off key 4-6 is pressed and held until 'UNIT' is shown on the display and then 

the zero key 4-3 is pressed.  The measuring unit changes over to the other one when 
the on/off key is released again.  

 This process takes a total of 7 seconds (from pressing the zero or on/off button). 
 
g) To change to the measuring mode from 'simple' to 'continuous' or vice versa, the 
zero or on/off key 4-6 must be pressed and held down until 'SC' appears in the display. 
Then the zero key 4-3 is pressed. The symbol 

((-)) 

stands for the continuous measuring 

mode  and  'S'  for  the  simple  measuring  mode.  This  process  takes  9  seconds  (from 
pressing the on/off button).                   

6. 

Zero adjustment 

 
a) The zero adjustment for 'Fe' (=F) and  'NFe' (=N) should be done separately. The 
iron zero plate is taken when 'Fe' is read on the display, while the aluminium zero plate 
is taken when 'NFe' is read.  
The measuring probe 4-1 is now carefully placed on the zero plate. The zero/on/off key 
4-6 is pressed and '0' is shown in the display without lifting the probe. 

Attention: The zero adjustment is useless if the measuring probe is not directly 
on the zero plate or another, uncoated carrier material.  

b) A suitable adjustment foil, which corresponds approximately to the measuring range, 
must be selected. 
(c)  This  standard  alignment  foil  is  positioned  on  the  supplied  zero  plate  or  on  the 
appropriate uncoated substrate for the forthcoming tests. 
d) Carefully press the measuring probe 4-1 in the middle of the standard foil and then 
lift it off again. The reading on the display corresponds to the measured value of the 
layer thickness.  
This reading can be corrected by pressing the plus key 4-4 or the minus key 4-5. This 
should  be  done  if  the  probe  is  not  in  the  nearer  vicinity  of  the  zero  plate  or  the 
corresponding carrier material. 
e) The procedure described in d) should be repeated until the result is correct. 

7. 

Battery replacement 

 

Содержание TC 1250-0.1F

Страница 1: ...2336 Balingen e mail info kern sohn com Phone 49 0 7433 9933 0 Fax 49 0 7433 9933 149 Internet www sauter eu Instruction Manual Digital Coating Thickness Gauge SAUTER TC Version 2 0 04 2020 GB TC BA e 2020 PROFESSIONAL MEASURING ...

Страница 2: ...lity of this measuring device and its extensive functionality For any questions wishes and suggestions please do not hesitate to contact us Table of contents 1 Introduction 3 1 1 General description 3 2 Functions 3 3 Technical data 3 4 Description of the control panel 5 5 Measuring procedure 5 6 Zero adjustment 6 7 Battery replacement 6 8 Adjusting foils 7 9 General notes 7 10 Restore factory sett...

Страница 3: ... etc This probe is often used to measure the thickness of galvanized layers lacquer layers enamelled porcelain layers phosphorescent layers copper plates aluminium plates alloys paper etc The N probe measures the thickness of non magnetic layers on non magnetic metals It is used to measure anodizations lacquer layers glazes colours enamel plastic layers powder coating etc These should be on non ma...

Страница 4: ...ries UM 4 Ambient temperature 0 C to 50 C Air humidity 80 Dimensions 126 x 65 x 27mm Weight Ca 81g without batteries Scope of delivery Carrying case operating instructions Depending on the type 1 built in measuring probe F TC 1250 0 1 F with F N key 2 built in measuring probe N TC 1250 0 1 N with F N key 3 built in measuring probe FN TC 1250 0 1 FN without F N u S C key Spacer foils for each model...

Страница 5: ...tself by the symbol Fe F for ferrous metals or NFe N for non ferrous metals shown on the display It switches to the automatic measuring mode which also correctly recognizes and assigns the zero plate or any other substrate itself b Place the measuring probe 4 1 on the layer to be measured The layer thickness can now be read on the display The reading result can be corrected by pressing the plus ke...

Страница 6: ...ressing the on off button 6 Zero adjustment a The zero adjustment for Fe F and NFe N should be done separately The iron zero plate is taken when Fe is read on the display while the aluminium zero plate is taken when NFe is read The measuring probe 4 1 is now carefully placed on the zero plate The zero on off key 4 6 is pressed and 0 is shown in the display without lifting the probe Attention The z...

Страница 7: ...probe should only be carried out by qualified personnel can be carried out 10 Restore factory settings a It is recommended that the factory settings be performed in the following cases no more measurements can be made at all the measuring accuracy is impaired due to drastic changes in the environment in which the measuring instrument is used After replacing the measuring probe b Procedure The fact...

Страница 8: ...ng the on off button It takes about 11 seconds from the beginning of pressing the on off button Its value can be changed by pressing the Plus or Minus key after LN appears in the display and releasing the On Off key The value is stored and then the zero key is pressed A The reading result in the lower area is corrected by pressing the Plus or Minus key B The Ln value is increased if the reading re...

Отзывы: