4  Measuring Procedure 

 
4.1 The Power- key 3-6 has to be pressed to turn on the power. “0” appears on the 
display 3-2.                   

Note:

 The instrument will auto- calibrate, as soon as switched on. 

Please pay attention to the sensor. During the initial calibration, the sensor should be 
located neither in direct surroundings of the base plate nor other magnetic materials. 

 

4.2 The sensor has to be located onto the coating to be measured. 
The coating thickness will now be displayed.

 

 
4.3 To perform the next measurement, the sensor has to be lifted for more than 1cm     
off the base material and step 4.2 has to be repeated. 
             
4.4 In case of inaccuracies to the measurement result, it is recommended to 
calibrate the instrument before taking measurements, as described in chapter 6. 
       
4.5  The instrument can be switched off by the Power-  on/ Power-  off key 3-6. The  
gauge switches off automatically 2 minutes after the last key- operation. 

5  Offset-Accur 

 

With this instrument, you have got the possibility to improve the measurement result 
essentially by using the 

OFFSET-

Accur function. Therefore, it is necessary to cali-

brate the gauge with a reference coating in the typically measured range. This kind of 
fine adjustment can be also done with the calibration foils, included in delivery. Ide-
ally, this adjustment should be done on the base material which is actually used for 
the real measurement instead on the base plate, included in delivery. 
 
5.1 The result of the initially measurement is still shown on the display (as performed 
in 4.2). 
         
5.2. The reading on the display can be corrected by pressing the Plus- key or the Mi-
nus- key. During this operation, the sensor has to be removed from the base plate or 
the part going to be measured. 

6  Calibration (Adjustment) 

 

6.1 Zero calibration for „Fe“ and „NFe“ should be done separately. If you take the iron 
base plate, you will have to insert the F-sensor and „Fe“will be displayed. Or you take 
the base plate of aluminium; then you will have to insert the FN-sensor and "NFe"“will 
be displayed. 

Содержание TB

Страница 1: ... 72336 Balingen E Mail info sauter eu Tel 49 0 7433 9933 199 Fax 49 0 7433 9933 149 Internet www sauter eu Instruction Manual Digital Coating Thickness Gauge SAUTER TB Version 1 3 08 2017 GB TB BA e 1713 PROFESSIONAL MEASURING ...

Страница 2: ...nal range If you have any queries wishes or helpful suggestions do not hesitate to call our service number Models TB 1000 0 1 F TB 1000 0 1 N TB 1000 0 1 FN TB 2000 0 1F Summarize 1 Features 3 2 Specifications 3 3 Front panel description 4 4 Measuring Procedure 5 5 Offset Accur 5 6 Calibration Adjustment 5 7 Batteriewechsel 6 8 Adjustment foils 6 9 Correct handling at coating thickness measurement...

Страница 3: ...cs porcelain enamel copper zinc aluminium chrome lacquer layers galva nised or phosphorescent coatings alloy etc These layers should be located on magnetic metals e g steel iron nickel etc In N mode the coating thickness of nonmagnetic coatings can be measured which is located on nonmagnetic metals It is used to measure anodizing varnish paint enamel plastic coatings powder coatings etc These laye...

Страница 4: ...s valid within 50 μm around the Off Set Accur point Note All tolerances are valid after the adjustment procedure Smallest measuring surface 6mm Power supply 4x1 5 AA batteries included in standard delivery Operating conditions Temperatures 0 up to 50 C Humidity 80 Dimensions 161 x 69 x 32 mm Weight about 260 g including batteries Scope of delivery Transport case Instruction manual Instrument and s...

Страница 5: ... key operation 5 Offset Accur With this instrument you have got the possibility to improve the measurement result essentially by using the OFFSET Accur function Therefore it is necessary to cali brate the gauge with a reference coating in the typically measured range This kind of fine adjustment can be also done with the calibration foils included in delivery Ide ally this adjustment should be don...

Страница 6: ...p 6 4 has to be repeated until the measurement accuracy is achieved 7 Batteriewechsel 7 1 If the battery symbol appears on the display of if the battery voltage is less than 4 8V batteries should be replaced 7 2 For this battery cover 3 8 has to be removed and batteries must be taken off 7 3 Batteries 4x1 5V AA have to be installed correctly into the case 7 4 If the instrument is not in use for an...

Страница 7: ...ase plate included in delivery By doing this measurement accuracy will be better right from the start 10 2 The sensor will eventually be worn off Its life will generally depend on the num ber of measurements taken and also on the surface roughness of the coating Re placement of a sensor should only be done by trained and experienced staff 10 3 The sensors of coating thickness gauges should only be...

Отзывы: