4
TB-BA-e-1713
- Standard: 3% of the measured value or min. ± 2.5 µm
Is valid within a tolerance range of ± 100
μm around the individually measured range,
if a two-point calibration has been performed within this tolerance range.
For model TB 2000-0.1F accuracy is:
- Standard: 5% of the measured value or min. ± 2.5 µm
Is valid within a tolerance range of ± 100
μm around the individually measured range,
if a two-point calibration has been performed within this tolerance range.
-
Off-Set Accur
Mode: 1% of the measured value or min. ± 1.0 µm
It is valid within ± 50
μm around the
Off-Set Accur
point.
Note: All tolerances are valid after the adjustment procedure!
- Smallest measuring surface: 6mm
Power supply: 4x1,5 AA batteries, included in standard delivery
Operating conditions: Temperatures 0 up to 50 °C
Humidity
≤ 80 %
Dimensions: 161 x 69 x 32 mm
Weight: about 260 g (including batteries)
Scope of delivery: - Transport case
- Instruction manual
- Instrument and sensor(s)
- 1 set adjustment foils, available at every model
- Base plate (aluminium or steel or both at type FN)
3 Front panel description
3-1 Sensor
3-2 Display
3-3 Zero-key
3-4 Plus- key
3-5 Minus- key
3-6 Power- on/ Power- off key
3-7 µm/ mil key
3-8 Battery compartment/ cover