background image

TC_car -BA-e-1712 

 

 

play. The instrument enters the auto mode  which can automatically recognize the 
base plate or base itself. 
b) The sensor 3-1 has to be placed onto a coating layer to be measured. The reading 
on the display is the thickness of the coating layer.  
c) For the next measurement the sensor 3-1 has to be lifted for more than 1cm, “0” 
appears again and step b) has to be repeated.  
d)  In case of inaccuracies to the measurement result,  it is  recommended to adjust 
(calibrate) the instrument before measuring, as described in chapter 5. 
e) The instrument can be switched off by pressing the Power- on/ Power- off key 3-3. 
The power will switch off 80 seconds after the last key operation. 
f) The measurement unit can be indicated with `µm` or `mil`. To convert: 
The  Zero/ Power-  on/ off key 3-3 has to be pressed and not to be released until 
`UNIT` appears on the display. The measurement unit changes by releasing. 
 All in all,  this operation lasts about 6 seconds (from starting pressing the Zero-/ 
Power-on/ off key). 
g) To change the measuring mode from `single` to `continuous  or vice visa, the Zero/ 
Power- key 3-3 has to be pressed and not released until `SC` appears on the display. 
The measuring mode changes into the other one when the key is released. This lasts  
about 8 seconds. The symbol 

((•)) 

on the display

 

indicates the continuous mode. To 

change back to single mode, just press the Power-key again for 8 seconds. If 

((•)) 

has disappeared on the display, you are back in single measurement mode.

 

5  Adjustment  

 
a) Zero adjustment: 
Zero adjustment for `Fe` (=F) and `NFe` (=NF) should be carried out separately. 
The iron base plate has to be used if `Fe` is shown on the display. The base plate of 
aluminium has to be used if `NFe` is shown on the display.    
The sensor 3-1 has to be placed carefully onto the base plate and the Zero-/ Power 
key 3-3 has to be pressed without lifting the sensor. `0` appears on the display. 
 

Attention: The adjustment will be invalid, if the sensor is not directly placed on 
the base plate or another uncoated material. 

b) An appropriate adjustment  foil, according to the measurement range, has to be 
selected. 
c) The standard adjustment  foil has to be put onto the base plate or the uncoated 
base material for the outstanding tests. 
d) The sensor 3-1 has to be placed in the middle of the standard foil and then be 
lifted. The reading on the display is the measured value. The displayed reading can 
be corrected by pressing the UP- button A or the DOWN-button B, which are located 
in the battery compartment. This has to be done while the sensor is away from the 
base plate or the material to be measured. 
e) Step d) has to be repeated until the result is correct.  

Содержание SAUTER TC car

Страница 1: ...36 Balingen E Mail info sauter eu Tel 49 0 7433 9933 199 Fax 49 0 7433 9933 149 Internet www sauter eu Instruction Manual Digital Coating Thickness Gauge SAUTER TC car Version 1 2 08 2017 GB TC_car BA e 1712 PROFESSIONAL MEASURING ...

Страница 2: ... are pleased with your high quality Thickness Gauge with its big functional range If you have any queries wishes or helpful suggestions do not hesitate to call our service number Summarize 1 Features 3 2 Specifications 3 3 Front panel description 4 4 Measuring Procedure 4 5 Adjustment 5 6 Battery replacement 6 7 Adjustment foils 6 8 Trouble shooting 6 9 Restore Factory Settings 6 ...

Страница 3: ...s e g steel iron nickel etc It is often used to measure the thickness of galvanizing layer lacquer layer porcelain enamel layer phosphide layer copper tile aluminium tile some alloy tile paper etc The N sensor measures the thickness of non magnetic coatings on non magnetic metals It is used on anodizing varnish paint enamel plastic coatings powder etc It can be applied on non magnetic materials li...

Страница 4: ...ils Base plate iron Base plate aluminium Optional accessories Cable software for RS 232C ATC 01 3 Front panel description 3 1 Sensor F and N inbuilt 3 2 Display 3 3 Zero key Power on Power off key 3 4 Battery compartment cover 3 5 Jack for RS 232C interface 4 Measuring Procedure a The Power on Power off key 3 3 has to be pressed to switch on the instru ment 0 appears on the display 3 2 The instrum...

Страница 5: ... key is released This lasts about 8 seconds The symbol on the display indicates the continuous mode To change back to single mode just press the Power key again for 8 seconds If has disappeared on the display you are back in single measurement mode 5 Adjustment a Zero adjustment Zero adjustment for Fe F and NFe NF should be carried out separately The iron base plate has to be used if Fe is shown o...

Страница 6: ...e the coatings are 9 Restore Factory Settings a It is recommended to restore factory settings in following cases The instrument does not measure any more Measurement accuracy is degraded caused by drastic environmental changes b How to restore To restore factory settings includes both Fe setting and NFe setting Please note which symbol is shown on the display You can restore only one of them or bo...

Отзывы: