Instruction Manual 

FH 

 

FH_SO-BA-e-1217 

 

6

 

 

FH Adjustment Procedure 
 

1  Start the 

instrument 

Press ON/OFF  

The green 
light is on 

2  Enter the 

adjustment 
procedure 

Directly after pressing 
ON/OFF, press PEAK 
and PRINT together for 
more times and 

very 

short intervals

 until the 

red light is on.  

The lower 
red light is 
on 

3  Instrument 

Type 

Press SET  
Directly after pressing 
PEAK + PRINT 
 

XXX N is 
shown 

a  (Back to the 

normal  
mode ???) 

(If you are in the 
operation mode again, 
the instrument has to be 
shut off. Start again with 
1. Eventually pressing 
the buttons faster) 

 

4  Select the 

instrument 
type 

Press 

▼▲  

The 
correspondi
ng value to 
the instr. 
appears in 
the display 

5  Save 

instrument 
type 

Press SET  

 

6  Enter the 

adjustment 
sequence 

Press MEMORY 

The upper 
red light is 
on 

7  Choose the 

adjustment 
weight 

Press UNIT and 
with

▼▲ specify the 

calibration weight in 
Newton (X kg * 9,81) 
 

The weight 
is written in 
the display 

8  Save the 

adjustment 
weight 

Press SET and UNIT 
simultaneously 

 

9  Attach the 

weight to the 
instrument 
(by hanging) 

Press ZERO  

The 
instrument 
changes to 
operation 
mode. 
Adjustment 
procedure is 
finished 

 

 

Содержание FH-M

Страница 1: ...ER FH incl rechargeable battery Carrying Case Charger 5 pcs M3x 8 screws Important annotation By pressing the RESET key on the right side of housing see illustration on the left individual settings and memorised values can be re set or erased in example for a new start of the instrument after an operating error Data in mm Length of cable between display and sensor 2500 mm Measurement Terms Track c...

Страница 2: ...or ON press 1 sec UNIT Press short Select unit N kg or lb Press for 2 sec Display return ZERO Three functions Zeros the measuring result Tara function Cleans the peak value in Peak mode Saves a setting in SET mode SET 1 x Press Upper Limit HidT Hi Lt To change press or see section High Low Limit function 1 x Press Lower Limit LodT To change press or see section High Low Limit function 1 x Press Mi...

Страница 3: ...her limit This function allows efficient testing of OK NOT OK measurements of similar or identical testing objects A lower and an upper limit value can be defined The instrument compares the individual measuring results with these limit values and shows the OK or NOT OK result by green or red light diode and by sound To set these limit values please see the SET Menu in section Operating keys Measu...

Страница 4: ...mall quantities are removed or added to the material to be measured incorrect measuring results can be displayed due to the stability compensation inside of the instrument Example Slow draining off of liquid from a container suspended from the instrument Do not attach a continuous load This can damage the measuring unit as well as the parts relevant to safety Prevent jolts torsion and oscillation ...

Страница 5: ...ment over the maximum permitted weight Danger of breaking Always make sure that there are no living beings or materials below the load that could be injured or damaged The hanging electronic instruments from the SAUTER instrument are only suitable for hand held use or use in a test stand They are not suitable for hanging from a mechanical hook e g a crane hook Observe the information in the operat...

Страница 6: ... XXX N is shown a Back to the normal mode If you are in the operation mode again the instrument has to be shut off Start again with 1 Eventually pressing the buttons faster 4 Select the instrument type Press The correspondi ng value to the instr appears in the display 5 Save instrument type Press SET 6 Enter the adjustment sequence Press MEMORY The upper red light is on 7 Choose the adjustment wei...

Отзывы: