background image

 

Sauter GmbH

 

Ziegelei 1 
D-72336 Balingen 
E-Mail: [email protected]

 

Tel:  +49-[0]7433- 9933-0 
Fax: +49-[0]7433-9933-149 
Internet: www. kern-sohn.com 

 

Instruction Manual 

FH 

 

FH_SO-BA-e-1217 

 

3

 

 

1  x  Press:  Data  Transfer  [rS232]  to  PC 
(PC) or to the printer (Print) or (in version U 
5.1) to the test stand (stand)  

 

 

 
- PC: Pressing SET once: Saves the Settings 
- PRINT: Pressing SET twice: sending data to the printer 
- STAND: Pressing SET three times: sending a signal to  
  the Test Stand to stop its movement (in version U 5.1) 
 

Back Light: 

 

 

PEAK: 

 

 

Three functions 

Track mode (continuous measurement) 

Peak mode (capture of maximum values) 

Auto-Peak  mode,  same  as  Peak-mode,  only 
without the „Min limit peak save“ function  

 

MEMORY:  
 

Saves  the  peak  values  to  calculate  the  average 
value (please see section 

“Memory mode” 

 
DELETE Function  

 
Deletes  stored  peak  values  (only  in 

„Memory 

mode

“ active) 

 
 
PRINT:  
 

Sends the stored peak values to a PC or Printer  
(plea

se see section „Data Output”)

 

 
High / Low limit function 

  

 

 
LEDs to display OK / NOT OK Tests 
 
▼  

Lower than lower limit 

OK 

OK 

▲ 

Higher than higher limit 

 
This  function  allows  efficient  testing  of  OK  /  NOT  OK 
measurements of similar or identical testing objects.  
A  lower  and  an  upper  limit  value  can  be  defined.  The 
instrument compares the individual measuring results with 
these limit values and shows the OK or NOT OK result by 
green or red light diode and by sound.  

To  set  these  limit  values,  please  see  the  SET  Menu  in 
section „Operating keys“. 
 

Measurement (Track Modus) 

 

Display  (1)  shows  the  continuous  force  in  a  defined 
direction (6)  

 

  

To zero the display, press:  
 
 

 
 
Peak-Hold Function (Peak Mode) 

 
Please press: 
 
 

Auto-Peak-Hold-Function(Auto-Peak Mode) 

 
Please press: 
 

 
 

Min. limit peak save 

 

 

 
This  function  allows  to  eliminate  unwanted  „Pre-Peak 
values“ that are lower than the main peak value (Fp). The 
“Min  limit  peak  save”  value  (Fo)  takes  care,  that  these 
“Pre-Peak values” are not saved.  
 
The  „Min  limit  peak  save“  function  is  only  possible  in  
Peak-Mode. 
 
To  set  this  Min  limit  value,  please  see  the  SET  Menu  in 
section „Operating keys“. 

 

Memory  mode

  and  average  value  (from  up  to  10  peak 

values) 
 
Saving peak values in the instrument 
  

 

Activating the „AUTO PEAK Function“ by PEAK key 

 

 

Deactivating the „Average Function“ by MEMORY 
key 

 

 

Now, all peak values are stored automatically in the 
instrument 

 

 

 

 

 

 

 

 

HoldT 

rS232 

 

Содержание FH-M

Страница 1: ...ER FH incl rechargeable battery Carrying Case Charger 5 pcs M3x 8 screws Important annotation By pressing the RESET key on the right side of housing see illustration on the left individual settings and memorised values can be re set or erased in example for a new start of the instrument after an operating error Data in mm Length of cable between display and sensor 2500 mm Measurement Terms Track c...

Страница 2: ...or ON press 1 sec UNIT Press short Select unit N kg or lb Press for 2 sec Display return ZERO Three functions Zeros the measuring result Tara function Cleans the peak value in Peak mode Saves a setting in SET mode SET 1 x Press Upper Limit HidT Hi Lt To change press or see section High Low Limit function 1 x Press Lower Limit LodT To change press or see section High Low Limit function 1 x Press Mi...

Страница 3: ...her limit This function allows efficient testing of OK NOT OK measurements of similar or identical testing objects A lower and an upper limit value can be defined The instrument compares the individual measuring results with these limit values and shows the OK or NOT OK result by green or red light diode and by sound To set these limit values please see the SET Menu in section Operating keys Measu...

Страница 4: ...mall quantities are removed or added to the material to be measured incorrect measuring results can be displayed due to the stability compensation inside of the instrument Example Slow draining off of liquid from a container suspended from the instrument Do not attach a continuous load This can damage the measuring unit as well as the parts relevant to safety Prevent jolts torsion and oscillation ...

Страница 5: ...ment over the maximum permitted weight Danger of breaking Always make sure that there are no living beings or materials below the load that could be injured or damaged The hanging electronic instruments from the SAUTER instrument are only suitable for hand held use or use in a test stand They are not suitable for hanging from a mechanical hook e g a crane hook Observe the information in the operat...

Страница 6: ... XXX N is shown a Back to the normal mode If you are in the operation mode again the instrument has to be shut off Start again with 1 Eventually pressing the buttons faster 4 Select the instrument type Press The correspondi ng value to the instr appears in the display 5 Save instrument type Press SET 6 Enter the adjustment sequence Press MEMORY The upper red light is on 7 Choose the adjustment wei...

Отзывы: