background image

 

Introduction 

 
Thank you for purchasing the handy type pH meter Model SK-620PH.   
 

-

 

This  pH  meter  is  designed  to  measure  hydrogen  ion  concentration  (pH)  and  temperature  (liquid 
temperature). Do not use this instrument for any other purposes.   
This instrument has a voltage measurement feature, but do not measure voltages (electromotive force) 
when  a  sensor  probe  other  than  the  PHP  series  is  connected.  Doing  so  could  damage  the  electronic 
circuit of this instrument. 

-

 

Read this instruction manual before using the SK-620PH and keep it in a safe place for future references. 

 
 

Important safety instructions 

 

 

The SK-620PH is not explosion-proof. Never use it for flammable materials. 

 

DANGER 

RISK OF EXPLOSION 

 

-

 

Do not use the pH standard solutions for any purposes other than to calibrate pH meters. 
The  solutions  are  not  for  drinking.  Never  try  to  drink  it.  If  accidentally  ingested,  consult  a  doctor 
immediately.   

-

 

For any other inquiries, contact the store where you purchased the instrument or our service network. 

 

Overview 

 
The SK-620PH pH meter is easy-to-operate instrument for measuring hydrogen ion concentration (pH) and 
liquid temperature. 
The  meter  can  be  calibrated  at  three  points  (pH4,  pH7  and  pH10)  and  provides  reliable  measurements  by 
using the built-in automatic temperature compensation. 

 
 

Features 

 

 

Large LCD for easy reading 

The pH value and the temperature are indicated on a large LCD. 

 

 

Auto Power-Off 

The "auto power-off" function turns off the power automatically if the unit is not operated for 20 minutes. 
This conserves battery life if you forget to turn off the unit. 

 

 

Auto temperature compensation 

Once the PHP-31 sensor probe is connected to your SK-620PH, it will perform automatic temperature 
compensation to produce more accurate pH measurements. 

Содержание 6400-00

Страница 1: ...Instruction manual for No 6400 00 Handy type pH meter Model SK 620PH No 3 Nishi Fukudacho Kanda Chiyodaku Tokyo 101 0037 Japan URL http www sksato co jp english M 09...

Страница 2: ...pH meters The solutions are not for drinking Never try to drink it If accidentally ingested consult a doctor immediately For any other inquiries contact the store where you purchased the instrument or...

Страница 3: ...functions Main unit 1 Display Displays measurement readings or status of the unit 2 PWR key Press this key to turn on the unit To turn off the unit press the key again 3 CAL key Enters the calibration...

Страница 4: ...plug LCD 12 Display of pH or mV Displays pH values or mV values in the measurement mode Display the calibration values in the calibration mode 13 Display of temperature Displays temperature values in...

Страница 5: ...he SK 620PH be sure to observe the followings This is a precision instrument Be careful not to drop or knock it The operating temperature range for this instrument is 0 to 50 C and a relative humidity...

Страница 6: ...ndard solution before use Do not turn on the unit during this immersion time The sensor probe electrode of this unit may be damaged or its useful life may be shortened if it is used for liquids such a...

Страница 7: ...bottle which contains the phthalate pH standard solution Remove the protection bottle before measuring Reinstall the protection bottle after measurement 1 Removing the glass electrode protection bottl...

Страница 8: ...auto power off Note Auto power off is no longer cancelled once the unit is turned off Perform the auto power off canceling each time as necessary Calibration mode By performing the 3 point calibratio...

Страница 9: ...dicate that the calibration data has been stored in the unit memory Now the pH 4 calibration is complete 8 The unit automatically proceeds to the pH 7 calibration Wash the glass electrode with water d...

Страница 10: ...ature value will be displayed For details of the error codes refer to Troubleshooting Relation between pH value and temperature When performing calibration set the pH value according to the temperatur...

Страница 11: ...e factory 2 Inspection and calibration For an accurate pH measurement perform calibration using the pH standard solution If there is a deviation from the expected value adjust it before performing a p...

Страница 12: ...nsert the AC adapter plug into the AC adapter jack on the unit If an AC adapter other than the one provided by us is used choose one that meets the specifications described below Primary output Voltag...

Страница 13: ...th new battery Installing and replacing battery The power to the unit is turned off The auto power off is turned on Cancel the auto power off on the main unit Before measuring Error code pH display se...

Страница 14: ...with temperature Readings are abnormal The battery have run out Replace with new battery Installing and replacing battery The glass electrode of the sensor probe is dirty Wash the glass electrode Abou...

Страница 15: ...ower than 80 rh no condensing Battery Life About 20 hours continuous Materials Case ABS resin Electrode Glass Probe Epoxy resin Dimensions Main unit W 72 H 182 D 31 5mm Probe 12 x 135mm L cord length...

Страница 16: ......

Отзывы: