7
Table 3-1 RDM Options
(
∗
default setting)
Display Code Parameter Options
Description
UPdt
Display
update time
0.1 - 10.0 s
(0.5)*
Defines interval between
display updates
ScrL
Auto scroll
nonE*
2-15 s
Disables/enables auto scroll
on common measurements
display (main screen) and
defines scroll interval
rEtn
Auto return
to the main
screen
diS*, En
Disables/enables auto return
to the main screen after 30
seconds of uninterrupted
use
VoLt
Primary or
secondary
voltage view
Pri*, SEc
Defines primary or
secondary voltage to show.
Ph.P
Phase
powers
display
mode
diS*, En
Disables/enables display of
phase powers in common
measurements (main
screen)
Fund.
Fundamenta
l values
display
mode
diS*, En
Disables/enables display of
fundamental values in
common measurements
(main screen)
dAtE
Date format
ndY*
dnY
Ynd
Defines the date format in
the RTC display: d=day,
n=month, Y=year
Each date format character
is set separately.
Addr
Instrument
address
1*-247
diSP
5
ScrL
diSP
0.5
UPdt
diSP
ndY
dAtE
diSP
diS
Ph.P
diSP
diS
rEtn
diSP
Pri
VoLt
diSP
diS
Fund.
diSP
1
Addr