Operating Instructions YTM15MA
21
Operation
y
If an adjustment has been performed:
The result of the adjustment is
displayed .
The result of the calibration and
adjustment can be called up later at
any time via the menu Settings ->
Calibration/Adjustment -> Calibration
Report -> Adjustment of the Heating
or Heating 2-point Cal .
t
Switch off the temperature
measuring instrument .
t
Remove the measurement cable
connector from the temperature
measuring device .
t
Allow the device to cool down .
t
Open the hood .
t
CAUTION
Danger of injury
due to hot surfaces! Observe the
information in Chapter 4 .2,
Page 9 . Remove the adjustment
disk from the pan retainer using the
disk holder (removal tool) and leave
to cool on a temperature-resistant
pad .
y
The calibration and adjustment of
the heating module is now complete .
Calibration Finished
Date and time
2016-04-07 09:43
Adjustment
Yes
Target value 1
100°C
Actual value 1
95.50°C
Deviation 1
-4.50°C
Target value 2
180°C
Actual value 2
180.20°C
Содержание YTM15MA
Страница 1: ...1000026973 Operating Instructions YTM15MA Heating Adjustment Set...
Страница 2: ......