33
Data Output Parameters
(Codes
8 1
to
8 16
)
Data can be transferred to the output port depending on
the particular stability state of the weighing system (stabil-
ity parameter: stable readouts).
You can choose to have data transferred regardless of
stability or only when the readout is stable.
lf you opt to have data transferred only when the weighing
system has stabilized, an output command will remain
stored until the system has stabilized.
For the auto print setting (automatic data output), weight
data are transferred continuously the moment you turn on
the power. lf you have additionally selected the stability
parameter for this Option, only data for stable readouts
will be output. To stop and start automatic data transfer,
press the print key.
Each signal received by the RxD line will initiate data
transfer (external data request).
Interfacing Devices with the Scale (RS Interface)
Make sure that the interface port is electrically connected
to the scale housing. The interface cable supplied as an
optional accessory is shielded and must be installed so
that the interface end is electrically connected to the
housing. This electrical connection may result in interfer-
ence caused by ground loops or by transient currents if
you have grounded the housing or connected the protec-
tive grounding conductor for AC power. lf necessary,
connect an equipotential bonding conductor to the scale.