– Place the complete display module (501 / 502) in front of the display
housing (506),re-attach the grounding cable (^) in the housing (506), and
plug the connecting cable (505) back into the display PCB (508).
– Place the display module (501 / 502) into the display housing (506) and
fasten it in place; see Figure RC-F-57.TIF on page 30).
– Now the balance is ready to operate.
Exchanging the Digital PCB
To be able to exchange the PCB, you will need the following:
– the PSION server (CAS version 4.4) or
– the SARTOCAS program for personal computers
Important Note:
Before starting, you should first try to transfer the data from the old digital PCB
and store them. If this is not possible, the data record can be retrieved from
the International Technical Support in Goettingen using a PC.
– Open the balance housing (see page 23).
– Remove the connectors (B, C, D and E) from the digital PCB (204).
– Remove the locking screws (201.A) and the data interface (201).
– Unscrew and remove the slot (201.B) for the data interface (201).
– Remove the 3 screws (W, X and Y) from the brace (203) and the digital
PCB (204).
– Take the old digital PCB (204) from the intermediate cover plate (205).
– Transfer the shieldings (Z) located above the power socket on the compo-
nent side and the soldered side from the old PCB to the new PCB (204).
– Place the new PCB (204) on the intermediate cover plate (205), align
and fasten it back on using the 3 screws (W, X and Y).
– Replace the slot (201.B) for the data interface (201), fasten it and plug in
the data interface (201).
– Now insert the connectors (B, C, D and E) back into the sockets on the
digital PCB (204).
– Close the balance housing (see page 24).
Important Note:
You must now use the SARTOCAS program for personal computers or the
MC1 server to transfer the data record to the balance. Please refer to the
program description for the proper procedure.