12
Calibration/Adjustment
You can calibrate/adjust the scale by pressing the
U
key (Zero/tare) (3)�
Calibration weights: 10,000 g; resolution: ± 0�016 g�
Always allow approximately 30 minutes for the scale
to warm up after connecting it to AC power and
before performing calibration/adjustment�
Press the
U
key (Zero/tare) (3) for 2 seconds;
“10000” is displayed�
Release the key�
Center the calibration weight on the load plate (15)�
Adjustment is performed automatically�
Remove the calibration weight�