24
Internal Span Adjustment
All LP/LA balances are equipped with a built-in motorized calibration weight.
– Turn on the balance and allow it to warm up.
– Zero the balance, if necessary (press
t
).
– Press F6
F6
F6
F6
F6; press F1
F1
F1
F1
F1 to select “
int. calibr.
” or “
int. adjust
” and press
F6
F6
F6
F6
F6 to start the internal span adjustment.
– The built-in calibration weight is applied automatically, the weight is
stored, and the deviation from the last span adjustment is displayed.
Note:
Depending on the menu code setting (under 1 10 X), either calibration/
adjustment is made automatically (code 1 10 1: calibrate, then auto-adjust),
or the deviation from the last linearization is displayed, but no adjustment is
made until you press a key (code 1 10 2: calibrate, then manual adjust).
– If the error exceeds the tolerance limits, press F6
F6
F6
F6
F6 to make the correspond-
ing adjustment, or press F1
F1
F1
F1
F1 to cancel the adjustment procedure.
– The balance then returns to the weighing mode.
External Span Adjustment
There are two ways to perform external span adjustment:
1. with the Sartorius service software (PSION Server or SARTOCAS Service
Software for PCs) - if the adjustment error is greater than
±
2%, or
2. with the balance in the service mode, as described in this manual.
External Span Adjustment with Manual Input of Weight Values
Note:
You only need to input the calibration weight for span adjustment if:
1.
the available weight is not sufficiently accurate (at least one digit
greater than the display accuracy), or
2.
you are working with a weight other than the recommended weight
(e.g., with a heavier weight, close to the maximum capacity).
– Press
e
to turn on the balance.
– The balance performs a self-test.
– When
0.00
g is displayed (depending on balance model), enter the ser-
vice password “
202122
” and press
ö
.
– Press F1
F1
F1
F1
F1 to activate the INPUT mode.
– Press F2
F2
F2
F2
F2 to select the “
cal./adj. weight
” menu item.
– Enter the conventional weight value, including all decimal places, using
the numeric keys.
– Press F1
F1
F1
F1
F1 to store the adjustment and press F6
F6
F6
F6
F6 to exit the SETUP program.
La-030e.BMP
La-001e.BMP
La-031e.BMP, La-032e.BMP
La-016e.BMP
La-034e.BMP
Содержание Master Pro LA12000P
Страница 50: ...50 Spare part list LP LA LP LA Exploded View Diagram 1 LP_exp_1 WMF LA LP...
Страница 52: ...52 LP LA Exploded View Diagram 2 LP_exp_2 EPS...
Страница 54: ...54 LP LA Exploded View Diagram 3 LP_exp_3 EPS...
Страница 56: ...56 LP_exp_4 EPS LP LA Exploded View Diagram 4...
Страница 58: ...58 LP_exp1a WMF LP LA Exploded View Diagram 1a LP LA...
Страница 60: ...60 LP LA Exploded View Diagram 2a LP_exp2a WMF A 201 202 211 203 207 206 216 215 214 B 217 208 218...
Страница 62: ...62 LP_exp_3a WMF LP LA Exploded View Diagram 3a 301 302 308 309 305 306 307 304 305...