16
reproTEST
– Activate the service mode (see page 11).
– Press
v
(several times) to select the “
reprotest
” menu item.
This tests the reproducibility of the automatic application of the built-in calibra-
tion weight.
The internal weight is applied, measured and removed 6 times.
At the end of the routine, the reproducibility value is displayed.
High Resolution - Display * 10
– Activate the service mode (see page 14).
– Press
v
(several times) to select the “
disp x10
” menu item.
The balance switches to SuperRange resolution and an additional digit is dis-
played (for test purposes and initial verification).
Max
5200
g d=
0.01
g
0000.00
g
S
Reprotest selec.
Max
5200
g d=
0.01
g
0000.00
g
S
disp. x 10 selec.