104
Preparation
●
Turn on the scale: press
e
> The Sartorius logo is displayed; a self-test is performed
●
Select the “Differential weighing” application in the Setup menu: press
M
FC Models:
§
Select the Application menu:
App
soft key
§
Select
Application 2
: Press the
q
soft key and then the
O
soft key
FCA Models:
●
Select the
Application parameters
: press the
q
soft key 2 x,
then the
O
soft key once
●
Select
Application 1 (basic settings)
: press the
O
soft key
●
Select
Differential weighing
: press the
Q
or
q
soft key,
repeatedly, if necessary
●
Confirm
Differential weighing
: press the
O
soft key
Differential
Weighing sequence
1)
Individual weighing
weighing
Consecutive individual weighing
o Combined weighing
Serial weighing
Tare weighing
No
o Yes
Result with
None
decimal point
1 decimal place
o 2 decimal places
3 decimal places
4 decimal places
5 decimal places
6 decimal places
7 decimal places
Autosave
o Off
values
On; first value at stability
On; last value at stability
2)
On; value bet. 70 – 130%
3)
Minimum load
None
for autosave
o 10 digits
20 digits
50 digits
100 digits
200 digits
500 digits
1000 digits
Save statistics
o No
Yes
Generate printout
None
o Automatic after backweighing
Automatic after initial and backweigh.
Automatic after tare, initial and backweigh
Include sample ID
o No
in text line
Yes
Wg. seq. key
No
o Yes
Clear sample after
o No
res + ind. wgh.
Yes
Last residual weight
o No
(ashing)
Yes
1
) Setting can only be changed
when the application is first run
and when the
Wg. seq
key
option is set to “No”
2
) The last value with the stability
symbol is saved only during
initial sample weighing. Tare
and backweights are saved as
the “first value at stability”. This
menu option enables you to
perform filling functions during
initial weighing.
3
) To autosave a value between
70 and 130% of the
initialization value, the scale
must be unloaded to below
30% or loaded to above 170%
of this initialization value.
* = Setting can only be changed when the programm is initially run
and when the
Wg. seq.
key option is set to “No”
o = factory setting
Содержание Factory Series
Страница 172: ...172 ...
Страница 185: ...185 Overview Dimensions Scale Drawings FC06BBE SX CE FC6CCE HX CE FC2CCE SX CE All dimensions are given in millimeters ...
Страница 193: ...193 Overview ...
Страница 194: ...194 ...
Страница 195: ...195 Overview ...
Страница 196: ...196 ...
Страница 197: ...197 Overview ...
Страница 198: ...198 ...
Страница 199: ...199 Overview ...
Страница 200: ...200 ...
Страница 201: ...201 Overview ...
Страница 202: ...202 ...
Страница 203: ...203 Overview ...
Страница 204: ...204 ...
Страница 205: ...205 Overview ...
Страница 206: ...206 ...
Страница 207: ...207 Overview ...
Страница 208: ...208 ...
Страница 209: ...209 Overview ...
Страница 210: ...210 ...