31
Operating the Combics
Weighing
W
The basic weighing function is always
accessible and can be used alone or in
combination with application programs,
such as Counting, Checkweighing,
Weighing in Percent, etc.
Features
– Zero the weighing platform by pressing
(
– Store the weight on the platform as tare
by pressing
)
– Enter a tare weight using the numeric
keys (press
)
to save)
– Enter a tare weight using a bar code
scanner
– Automatic taring of container weight
– Deleting tare values by entering
0
(press
)
to save)
– Press
k
to toggle the display
between:
– Gross and net values, or
– 1st and 2nd weight unit, or
– Normal and 10-times higher
resolution
Define the function of the
k
key in
Setup, under:
Fn-key
– Weigh with two weighing platforms
– Individual ID codes for weight values
– Print weight values:
– Manually, by pressing
p
– Automatically
(see “Data Output Functions”)
– GMP-compliant printout
(see “Data Output Functions”)
– Restore factory default settings.
Configure in Setup under:
Application: Weigh:
Fty settings
Soft Key Functions
ID
Enter up to four ID codes
for identifying results on
the printout
ID 1
Save the value entered as
the first ID code
Preparation
§
Activate the Setup program:
Press the
M
key
§
Select Application Parameters:
Press the
O
soft key
§
Select “Weighing Only”:
Press the
O
soft key
Weigh only
Minimum load for autotaring
1 digit
2 digits
5 digits
o 10 digits
20 digits
50 digits
100 digits
200 digits
500 digits
1000 digits
Autotare first weight
o Off
On
Factory settings
Yes
o No
o = factory setting
§
Save settings and exit Setup: Press the
M
key or the
oo
soft key
Automatic Taring
The first weight on the scale that
exceeds the preset minimum load is
stored in the tare memory at stability.
The values for subsequent loads are
stored as weight values. The scale
returns to the initial state when the
load is less than 50% of the minimum
load.
Configure in Setup under:
Application: Weigh:
Autotare 1st weight
Minimum Load
The minimum load required for
automatic taring of the container
weight on the platform (first weight),
or for automatic printout of results,
is configured in Setup under:
Application: Weigh:
MinL. autotar
You can choose from the following
10 levels for this setting:
1 digit (no minimum load)
2 digits
5 digits
10 digits
20 digits
50 digits
100 digits
200 digits
500 digits
1000 digits
The “digits” here refer to the scale
intervals in the connected weighing
platform. If the interval of the
connected platform is 1 g, for example,
and 1000 digits are required, you must
place at least 1000 g (= 1000 intervals
= 1000 digits) on the weighing platform
for autotaring.
Automatic Printing
The first weight value that exceeds
the minimum load is printed out
automatically. Configure in Setup
under:
Device: Printout: Once
at stability
Weighing with Two Weighing
Platforms
You can connect two weighing
platforms to the Combics 3. Press the
n
key to toggle the display between
weighing platforms.
You can define which of the two
platforms is active in the display when
the Combics is switched on. This is
configured in Setup, under:
Device: Operat.: Main
scale
The display shows the readout from
the main scale when you switch on the
Combics.
Содержание CH3E
Страница 73: ...Dimensions for CH Platforms Scale Drawings CH3G CH3E 73...
Страница 80: ...80...
Страница 81: ...81...
Страница 82: ...82...
Страница 83: ...83...
Страница 84: ...84...
Страница 85: ...85...
Страница 86: ...86...
Страница 87: ...87...
Страница 88: ...88...
Страница 89: ...89...
Страница 90: ...90...
Страница 92: ...92...
Страница 94: ......
Страница 95: ......