Sartobind

®

S Demo Kit

Purification of a protein with Sartobind S Membrane Adsorbers

85034-535-14

1. Storage before use

The kit solutions have been filtered
sterile (0.2 μm) but no preservatives
have been added to the buffers. Thus
the kit should be kept refrigerated at
+4 to +8°C after opening bottles.

2. Introduction

The kit has been developed to
demonstrate the capability of
Sartobind Membrane Adsorbers to
purify quickly and easily a protein
(cytochrome c) from a contaminant
(Amido Black).

A trained person can demonstrate
the purification in one minute. 
A beginner can perform the purifica-
tion within five minutes.

3. Kit contents

Sartobind S 15

1

Equilibration 
buffer

+

50 ml

Loading solution 50 ml

Elution buffer

50 ml

10 ml syringe

1

Order code 

90-KIT-01

4. Technical data Sartobind S 15

Membrane volume 0.41 ml

Bed height

0.8 mm

Number of layers 3

Membrane 
diameter 25 

mm

Dimensions

36 

+

28 mm

Ligand

Sulfonic acid

Degree of 
substitution

2 – 5 μeq/cm

2

Binding capacity 0.8 mg/cm

2

29 mg/ml
lysozyme

Membrane matrix Cross-linked rein-

forced cellulose

Nom. Pore size

>3 μm

Membrane 
thickness 275 

μm

Spacer

No spacer

Membrane area

15 cm

2

Housing

Polysulfone

Connector inlet

Female Luer 

outlet Male Luer 

Flow at 0.1 MPa 
(1 bar, 14.5 psi)

130 ml/min

Recommended 
flow rate

10 –150 ml/min

Maximum 0.6 

MPa 

pressure

(6 bar, 87 psi)

Chemical 

All commonly used 

stability

chromatography
buffers

pH stability
Long term 

2–12

Short term

1–14

Storage

in equilibration
buffer containing
20 % ethanol

4. Operation of the units

We recommend to use the enclosed
syringe for the demonstration: 
Use recommended volumes  as listed
in table.

Buffers, protein solution and volumes

Equilibration 

10 mM potassium 

buffer: 10 ml

phosphate buffer 
pH 7.0

Loading solution:  Cytochrome c 
5 ml   

(2 mg/ml) in 
equilibration
Amido
Black (5 μg/ml)

Washing buffer:  Equal to  
5 –10 ml

equilibration buffer

Elution buffer: 

Equilibration 

3 ml 

1 M NaCl

The buffers and protein solution can
be prepared by yourself to repeat the
demonstration. S 15 unit is reusable.

4.1 Equilibration

Hold the unit in a vertical position.
Remove the inlet cap. Fill 10 ml
syringe with 10 ml equilibration
buffer and connect to the inlet of the
unit. Remove the outlet cap. Fill the
unit and remove any remaining air by
moving the plunger up and down
with short strokes. Make sure that
the unit is completely filled with the
fluid. 

Important note:

The unit has to be filled completely
with fluid throughout the operation
to ensure even flow and distribution
of the feed stream.

Outlet

Inlet

Flow
direction