7.9 Draining
You may drain the capsule by application of air or nitrogen pressure
(<1 bar |14.5 psi) to the inlet of the capsule.
!
Important
A dual air regulator system is recommended to prevent over-pressur-
ization of the Sartobind capsule. The first regulator should reduce
line air pressure to the Sartobind capsule to 2 bar. The second
regulator, positioned immediately upstream of the Sartobind
capsule, should reduce the 2 bar regulated supply pressure to the
<1 bar (14.5 psi) draining pressure.
7.10 Regeneration and storage
After use, regenerate with e.g. 50% ethylene glycol, 70% ethanol
or 30% isopropanol in pure water, wash extensively with pure water
and 20% ethanol and store airtight in 20% ethanol at +4 to +8°C
in a dark place. Do not store in high salt solution.