Introduction

Introduction

Introduction

Introduction

    

The BioWelder® is working with a 
thermal welding cycle that does not 
require the use of a laminar flow.  
In this welding process the blade  
is kept sterile and endotoxin free 
throughout the whole process  
in accordance to the reaction 
equation and the Arrheniussatz 
through the use of constant time 
and temperature (Wallhäuser, the 
D-Value resistence indicator for 
bacillus subtilis, is less than 1 sec 
at 200°C dry heat). The inserted 
hoses will be cut by a blade and 
then fused together through the 
unit. The total process will take 
place under sterile conditions.  
The cycling process is completely 
automated and will only be 
released after a complete cycle.  
Different programmable holders 
make it possible to weld 
thermoplastic hoses of the same 
type together(such as C-Flex®, 
Sanipure®, Pharmed®) with 
diameters of 6.4 mm (1/4”) up to 
19.0 mm (3/4”). 
 
 

    

2.1

2.1

2.1

2.1    

Features

Features

Features

Features    

 

–  Fully automated operation 

 

–  Welds thermoplastic tubing up to 

3/4 inch O.D 

 

–  Preinstalled settings for C-Flex®, 

Raumedic® FluiSoft

TM

, AdvantaPure 

AdvantaFlex®, Pharmed® and 
Sanipure® tubing 

 

–  True thermal weld provides  

exceptional strength 

 

–  Aseptic connection without  

a laminar flow hood 

 

–  Data output by a printer 

 

–  Compact design, portable 

 

–  Easy to validate, with validation 

equipment 

 

–  Can weld many different tube 

diameters 

 

    

2.2

2.2

2.2

2.2    

Safety

Safety

Safety

Safety

    

The device may be used  
by children over the age of  
8 years of age and by people 
with reduced physical, 
sensory or mental 
capabilities, but only after 
they have been instructed 
about how to use the device 
safely in order to understand 
the associated hazards. 
Children must not be 
allowed to play with the 
appliance. Cleaning and 
maintenance of the 
equipment should not be 
done by children without 
adult supervision. 
 

The blade reaches over 400°C 
during the welding cycle! To avoid 
possible C burns DO NOT open the 
access door unless it is released. 
There will be an automatical 
release of the cover as soon as the 
blade cooled down sufficiently. 
Make sure that the unit is 
unplugged before performing  
any maintenance or adjustments.  
Use only tubing that is compatible 
with the unit. Please consult 
Sartorius Stedim Biotech for a list 
of tubing types that have been 
tested. Only use BioWelder® cutting 
blades. For optimal operation, do 
not reuse blades. Do not attempt 
to connect liquid-filled tubing, or 
drop liquid into the unit. 
 

    

2.3

2.3

2.3

2.3    

Restrictions

Restrictions

Restrictions

Restrictions    

The BioWelder® has been tested to 
produce welds of sufficient quality 
and strength to permit sterile  
transfers.  
 
Example validation protocols and 
test results are provided for 
reference. However, it is the 
responsibility of the end user to 
verify and validate the suitability  
of this device for their application. 
Under no circumstances can data  
in this manual be submitted to any 
regulatory agency in support of a 
clinical product without specific 
written permission from the 
manufacturer. 
 
The connection made by the  
BioWelder® may not be satisfactory 
for your application. It is the user’s 
responsibility to determine the  
suitability of the machine. 
Connections should be carefully 
inspected before use, as even the 
best equipment always has the 
possibility of mechanical or 
electrical failure. 
 
The BioWelder® has been  
programmed for use in room  
temperature (20-30°C) 
 
The Biowelder unit can be 
programmed with maximum  
3 parameter sets. When ordered 
without any mention, the 
Biowelder comes with C-Flex®, 
PHARMED® BPT and Sanipure® 60. 
The unit can be ordered with other 
parameter sets. In that case, please 
specify the parameter sets you 
want to have in the unit and the 
ones you want to have removed. 

 

Содержание BioWelder

Страница 1: ...Operation Manual Operation Manual Operation Manual Operation Manual BioWelder 85030 531 87...

Страница 2: ...dix 4 Short form of the manual for the calibration holder 21 1 1 1 1 Manual Manual Manual Manual Please read this manual carefully It includes important information for the safe and successful use of...

Страница 3: ...e of the equipment should not be done by children without adult supervision The blade reaches over 400 C during the welding cycle To avoid possible C burns DO NOT open the access door unless it is rel...

Страница 4: ...mains sterile Once the tubing has been cut the machine slides the cut tubing pieces into alignment The blade is then extracted while the tubing is pushed together The tubing pieces remain at all times...

Страница 5: ...blade Step 2 Blade heats up sterilization and then cuts through both tubing pieces Step 3 Tubing holders slide the cut tubing into alignment Hot blade is held against the cut tubing to ensure sterilit...

Страница 6: ...upper and lower tubes are sealed on both sides Step 2 Blade heats up sterilization and then cuts through four tubing pieces Step 3 Tubing holders slide the cut tubing into alignment Hot blade is held...

Страница 7: ...anguage Choose the language Choose the language Choose the language You can choose the language of your unit German or English To open this menu press MODE and the right arrow Enter the password 264 U...

Страница 8: ...s you want to connect with a suitable disinfectant Place the hoses carefully in the tube holders The hose ends have to extend beyond the tube holders Attention Attention Attention Attention Check that...

Страница 9: ...old the middle of the blade between two fingers Insert the blade into the blade holder and check that the blade locks in correct position and the point shown in the left picture shows to the temperatu...

Страница 10: ...s The tubing holders can be quickly changed to connect different diameter tubing The holders must be changed as a set Left and Right holders Removing the holders 1 Open the cover and both tube holders...

Страница 11: ...ot take care about the following explanations there is the risk of a damaged hose which can cause a leackage Step 1 Press the hose on the back side down to the bottom with both tumbs Check the hose po...

Страница 12: ...rect position and has to be reinserted Check again the correct position of both hoses Step 5 Press both clamps sufficently down to grip the tubes Step 6 Finally close the blue clamp If both hoses are...

Страница 13: ...erent types of hoses types of hoses types of hoses types of hoses It is possible to weld thermoplastic tubings like C Flex Raumedic FluiSoftTM Pharmed and Sanipure It is not possible to weld non therm...

Страница 14: ...nation temperature welding temperature welding time and the date and time 7 3 7 3 7 3 7 3 Printing of last 10 welding protocols Printing of last 10 welding protocols Printing of last 10 welding protoc...

Страница 15: ...installed Check that the surface of contact plate blue holder and the contact pins in the BioWelder are clean Depyrogenation temperature can not be reached Has the blade been changed in the right cyc...

Страница 16: ...e limitation heater Blade is not able to reach the depyrogenation temperature in between the programmed 90 sec Legend Z axis corresponds to the stroke X axis corresponds to the red holder Y axis corre...

Страница 17: ...irs must be done by a factory service representative Exception is the cover which can be exchanged by the client itself 9 3 9 3 9 3 9 3 Calibrat Calibrat Calibrat Calibration ion ion ion The unit has...

Страница 18: ...are available as standards Other sizes can be provided upon The maximum tubing O D is determined by the cutting blade width This is currently restricted to 3 4 inch O D max OD OD OD OD ID ID ID ID Wa...

Страница 19: ...ing Some of the tubing brands that have been tested are listed below The unit cannot be used to join silicone tubing or PTFE Pharmed Ideal very high purity tubing Can be autoclaved or SIP Long life in...

Страница 20: ...down arrow English and German 6 As soon as you have chosen the language press MODE 7 The BioWelder is now ready for welding Set date and time Set date and time Set date and time Set date and time 1 P...

Страница 21: ...t form of the manual for the calibration holder Short form of the manual for the calibration holder Clean the Infrared sensor with a cotton stick and alcohol Choose the program Temp validating in the...

Страница 22: ...re indicated in the display Insert a new blade in the way displayed Insert the Calibration Holder very carefully do not touch the thermocouple and close the holder afterwards Measuring instrument has...

Страница 23: ...t of range is where you can see a high temperature decrease During this phase you don t see big temperature jumps Afterwards the unit cools down the blade sufficiently to a safe temperature for the op...

Страница 24: ...the prior written permission of Sartorius Stedim Biotech GmbH The status of the information specifications and illustrations in this manual is indicated by the date given below Sartorius Stedim Biote...

Отзывы: