NOTE!
When “cellulose 0.8 μm” is selected, the air flow rate is limited to 6 m
3
/h.
If you have selected an air flow rate setting higher than 6 m
3
/h and then enter the filter class
parameter “cellulose 0.8 μm”, the air flow rate will automatically be reset to 6 m
3
/h.
Other filter types than those listed can be used, but you must then check the regulatory
behaviour and the air throughput of the MD8 airscan by tests/recalibration.
When a displayed parameter is changed with the
⇑
or
⇓
key, the new setting flashes to show
it has not been stored. If you wish to store the change, make sure to press
Enter
. If you exit
the parameter input menu, either by pressing the
Stop
or
Parameter
key, without first
pressing the
Enter
key, all changes will be ignored.
If you exit a menu item by pressing
Stop
, the current function (parameter display or input,
service menu or calibration) will be aborted, and the instrument returns to the initial
starting mode.
You can return to the initial starting position at any time by pressing
Stop
.
24 |
Содержание MD8 airscan
Страница 1: ...85030 511 25 Operating Instructions Air Sampler MD8 airscan Ver 05 2011...
Страница 2: ...Read these operating instructions carefully before switching the instrument on 2...
Страница 18: ...Fix a gelatin filter unit to the adapter Fig 8 In principle sampling can now be started 18...
Страница 36: ...Fig 14 Fig 15 36...
Страница 37: ...Fig 16 Fig 17 37...
Страница 38: ...Fig 18 38...
Страница 40: ...Fig 19 40...
Страница 52: ...17655 17656 17657 17657 17657 17208 52...
Страница 53: ...17085 17088 17662 16970 16970 16970 53...
Страница 54: ...17801 17016 17030 17033 5185307T9 SS A 54...