69
36
BACK-UP SYSTEM (OPTION)
With the back-up system (CVK-UB4), the freezer prevents the chamber temperature from going up by
injecting the liquid CO
2
gas when the power supply is disconnected (power failure, disconnection of power
cord, breaker OFF) or in the case of failure of freezer itself. The liquid CO
2
gas is injected with the
activation of solenoid valve energized by battery when the chamber temperature reaches the alarm
temperature.
Following shows the procedure for setting the back-up system.
1.
Setting of liquid CO
2
gas cylinder
By using the joint and pipe enclosed with the
back-up system, connect the liquid CO
2
gas
cylinder to the joint of the back-up system. For
this setting, consult with a qualified gas supplier or
Sanyo sales agency.
2.
After setting the liquid CO
2
gas cylinder, operate
the freezer until the chamber temperature reaches
the required level.
3.
Set the temperature of back-up operation at the
temperature higher than -70
o
C. The back-up
system is operated continuously if the temperature
of back-up operation is set at the temperature
lower than -70
o
C. This means the liquid CO
2
gas
is consumed very quickly.
4.
Switch on the back-up system.
5.
Make sure that liquid CO
2
gas spouts into the freezer chamber by pressing the back-up test switch.
Note:
Ŷ
The liquid CO
2
gas cylinder loses its cooling capacity at speed when the ambient temperature is over
31
o
C. Install the liquid CO
2
gas cylinder in the cool environment. And the duration of back-up time per
one liquid CO
2
gas cylinder varies depending on the ambient temperature. Refer to “Installation of
back-up system” enclosed with the back-up system for the available back-up time.
Ŷ
The liquid CO
2
gas cylinder should be a siphon type.
Ŷ
Use the pipe encloses with the back-up system for the setting. (The extension of the pipe is not
permitted because of cooling capacity.)
Rough scheme (back side)
Liquid CO
2
gas cylinder
Содержание MDF-C8V
Страница 8: ...5 Dimensions 405 490 R 2 5 550 58 852 945 425 887 35 727 685 33 㧨 8㧪 ...
Страница 9: ...6 Refrigeration circuit diagram ...
Страница 12: ...9 Components on PCB ...
Страница 15: ...12 Wiring diagram ...
Страница 16: ...13 Circuit diagram ...
Страница 36: ...33 MDF C8V Ultra Low Temperature Freezer INSTRUCTION MANUAL ...
Страница 44: ...41 8 FREEZER COMPONENTS 13 12 5 11 1 2 3 6 7 8 5 4 10 9 ...
Страница 76: ...73 CVK UB4 CVK UB4 I Back up System for Ultra Low Temperature Freezer INSTRUCTION MANUAL ...
Страница 92: ...1 JK1659 150B ...