47
14
CHAMBER TEMPERATURE SETTING
Table 1 shows the basic procedure for setting the chamber temperature. Perform key operations in the
sequence indicated in the table. The example in the table is based on the assumption that the desired
temperature is -75
o
C.
Note:
The chamber temperature is set to -80
o
C at the factory.
Table 1 Basic operation sequence (Example: Chamber temperature -75
o
C)
Description of operation
Key operated
Indication after operation
1
Connect the power cord to the
dedicated outlet.
----
The current chamber temperature is
displayed.
2 Press SET key.
SET
The second digit is flashed.
When pressed, the figure of settable
digit changes.
3
Set to -75 with the numerical value
shift key and digit shift key.
When pressed, the settable digit
is shifted.
4 Press SET key.
SET
Set temperature is memorized and the
current chamber temperature is
displayed.
Note:
Ŷ
The temperature set mode returns to the temperature display mode automatically when 90 seconds has
passed without any key operation.
Ŷ
Although the value of the chamber temperature setting can range from -55
o
C to -85
o
C, the guaranteed
temperature range with no load is between -60
o
C and -80
o
C when the ambient temperature is 30
o
C.
KEY LOCK FUNCTION
This unit is provided with the key lock function. When the key lock is ON, change of temperature setting
through the key pad is not available. The key lock is set to OFF at the factory.
Display Mode
Function
L 0
Key lock is OFF
Enable to change of temperature setting
L 1
Key lock is ON
Disable to change of temperature setting
Table 2. Procedure for key lock setting (change from key lock OFF to key lock ON)
Description of operation
Key operated
Indication after operation
1
----
The current chamber temperature is
displayed.
2 Press digit shift key for 5 seconds.
The first digit is flashed.
3
Press numerical value shift key
and scroll the figure to 1.
When pressed, the figure of settable
digit changes.
4 Press SET key.
SET
The key lock is set to ON.
The current chamber temperature is
displayed.
Содержание MDF-C8V
Страница 8: ...5 Dimensions 405 490 R 2 5 550 58 852 945 425 887 35 727 685 33 㧨 8㧪 ...
Страница 9: ...6 Refrigeration circuit diagram ...
Страница 12: ...9 Components on PCB ...
Страница 15: ...12 Wiring diagram ...
Страница 16: ...13 Circuit diagram ...
Страница 36: ...33 MDF C8V Ultra Low Temperature Freezer INSTRUCTION MANUAL ...
Страница 44: ...41 8 FREEZER COMPONENTS 13 12 5 11 1 2 3 6 7 8 5 4 10 9 ...
Страница 76: ...73 CVK UB4 CVK UB4 I Back up System for Ultra Low Temperature Freezer INSTRUCTION MANUAL ...
Страница 92: ...1 JK1659 150B ...