CALIBRATION
Temperature calibration
1.
Press the calibration key (CAL) for approximately 5 seconds again to enter the temperature calibration
input mode.
2.
The left digit of the digital temperature indicator blinks, and the digital CO
2
density indicator goes out.
3.
Set the present correct temperature with the digit shift key ( ) and numerical value shift key ( ) ,
then press the enter key (ENT).
4.
The incubator automatically reverts to the display mode.
[Example]
If the displayed chamber temperature is 37.0
o
C (set value) and the actual temperature is 36.8
o
C.
1.
Press the calibration key (CAL) for about 5 seconds.
2.
The “3” on the digital temperature indicator blinks, and the digital CO
2
density indicator goes out.
3.
Adjust the set value to the actual value of 36.8
o
C with the digit shift key ( ) and numerical value shift
key ( ), then press enter key (ENT).
4.
The incubator automatically reverts to the display mode.
31