- …27…-
○
Measurement unit setting
When the display shows [
], press the CH1 button or the
CH2 button to select its measurement unit. The selected unit is
indicated by the LED. The factory default setting is “m/min”.
○
Analog output full scale setting
When the display shows [
], press the CH1 button or the
CH2 button to select the maximum value when 2 V is output as
the analog output.
Measurement unit
Full scale /2 V
Factory setting
m/min
20/200/2000
2000
r/min
200/2000/20000
20000
mm
20/200/2000
2000
○
Averaging count setting
When the display shows [
], press the CH1 button or the
CH2 button to select the averaging count between 1 and 10
times. The factory default setting is 1.
5-3 Function Description
5-3-1 Maximum value hold function (MAX value)
– m/min & r/min measurements only
When the ENT button is pressed, the measurement unit LED flashes
and the CH2 display shows the maximum value that has been held.
Pressing the ENT button again cancels the maximum value display.
Measurement mode
Max. value
Flashing