Accuracy in the case of sine wave.
◎
Accuracy calculation
For example
:
Measurement DCmV
Displayed value
:
100.0mV
Accuracy
:
400mV range
・・・・±
(0.7%rdg+3dgt)
Error
:±
(100[mV]
×
0.7%+3[dgt] )=
±
1.0[mV]
True value
:
100.0[mV]
±
1.0[mV](In a range of 99.0
〜
101.0mV)
*
3[dgt] in the 400mV range corresponds to 0.3mV.
Specifications and external appearance of the product
described above may be revised for modification without
prior notice.
・
Buzzer sounds at less than 10
〜
120
Ω
・
Open voltage: Approx. 0.4V
Open voltage: Approx 1.5V
Function
Hz
DUTY
BUZZER
DIODE
Range
9.999Hz
99.99Hz
999.9Hz
9.999kHz
60.00kHz
0.1
〜
99%
Accuracy
±
(0.7%rdg+5dgt)
-------------------------------------
Remarks
*Accuracy in the case of sine
wave.
9.999Hz
〜
9.999kHz: 10Vrms
〜
250Vrms.
60.00kHz: 40Vrms
〜
100Vrms.
About input sensitivity and
frequency characteristic:
(Square wave DUTY 50% input)
2.5V 0 to peak input:
≧
1kHz
6V 0 to peak input:
≧
10kHz
40V 0 to peak input:
≧
60kHz