− 9 −
During the measurement, LED indicator lights and
symbol
turns on.
③
Press the TEST button again to stop the measurement.
④
Explanation when displaying values other than numbers
• In case of overload, the display is "O.L.".
• When the resistance of the auxiliary earth electrode H (C)
exceeds about 12 kΩ, the display is "- - - -".
• If there is no measuring current due to disconnection of E and
H (C), the display will be "- - - -".
• It will give unstable reading when S (P) is disconnection. If
you replace S (P) and H (C) on the terminal, you may notice
connection failure.
If the indication is "- - - -", rework to place auxiliary earth
electrode into the ground again (See 5-5-1). Also "
" may
be displayed
when turns on the display.
⑤
When the measurement is over, clean the auxiliary earth
electrode with wet cloth, etc. Please keep it drying to avoid
rusting.