Accuracy calculation method
Example) DC voltage measurement (DC mV)
True value: 100.0 mV
Range accuracy: 600 mV range - ±(0.5 %rdg + 2 dgt)
Tolerance: ±(100.0 [mV] x 0.5 % + 2 [dgt]) = ±0.7 [mV]
Displayed value: 100.0 [mV] ± 0.7 [mV] (within the range of 99.3
to 100.7 mV)
CAUTION:
If there is presence of strong magnetic field generated by
conductor fed large current, motors or strong radio wave, a
measurement may not be accurate.
Specifications and external appearance of the product
described above may be revised for modification without prior
notice.