Sanosil Q-Jet Advanced

 

 

 

 

 

 

 

2.5

 

Disclaimer

 

Sanosil declines any claims for damages or warranty claims if:

 

 

The product is used for any other purpose but the ones stated in this Operation Manual.

 

 

Regulations, standards or instructions in this Operation Manual are not adhered to.

 

 

The product is modified or transformed in any manner.

 

 

The product is not repaired by an authorized dealer and is used with other but the original spare parts provided by

 

Sanosil Ltd.

 

 

The product is used despite obvious safety faults or recognizable damages.

 

 

The product was exposed to mechanical tappets or has been dropped.

 

 

Accessories are used which are not supplied nor specifically authorized by Sanosil Ltd.

 

 
 

2.6 

Room Disinfecting with the Sanosil Q-Jet Fogger 

 

How to use the Sanosil Q-Jet Advanced: 
Fogging of Sanosil Disinfectants (only use Sanosil ready for use products containing up to 7.5% hydrogen Peroxide) 
can be used for disinfecting hard, nonporous surfaces in rooms as an adjunct following regular cleaning and 
disinfecting of room surfaces. Sanosil ready to use products are recommended for use in fogging operations as a 
supplemental measure after regular cleaning and disinfecting procedures. Pre-cleaning protocol defined by 
healthcare facility. If other cleaning/disinfecting products then Sanosil products are used, surfaces MUST BE DRY 
before fogging. The more thorough the surfaces are cleaned beforehand, the more effective the adjunct 
disinfection treatment will be. 
 
Areas should be cleaned daily, after a contamination has been detected, or on an as needed basis depending on 
the volume of traffic and use. 

 

  Users should familiarize themselves with Q-Jet operation by reading the Q-Jet User Manual prior to first use. 

It is essential that Q-Jet users follow all safety warnings as detailed in this operating manual. 

 

It is recommended that users read the entire document before proceeding with room disinfecting treatment. 

 
 

2.7 

Different ways to use the Sanosil Q-Jet Advanced 

 

a.  Overnight disinfection (recommended): 

Procedure: 

Disinfecting process according procedure on page 15, with extended exposure time. 

Advantages: 

Extended exposure time, maximal disinfection efficacy, hydrogen Peroxide level will lower  
automatically overnight and has to be measured only once on the next day

 

Disadvantages: Room cannot be used and Sanosil Q-Jet Advanced is not available overnight 

 

b.  Standard disinfection without additional measures: 

Procedure: 

Disinfecting process according procedure on page 15 with standard waiting time. 

Advantages: 

Time consumption: 3-10 hours (depends on the room size, humidity and the decomposition 

 

time of hydrogen peroxide in the treated room), Sanosil Q-Jet Advanced and room can be re-used 

faster 
Disadvantages: Additional personal and time needed for hydrogen peroxide level measurements 
 

c.  Standard disinfection with speed up measures: 

Procedure: 

Disinfecting process according procedure on page 15 with reduced waiting time. 

Advantages: 

Time consumption: 3-5 hours (depends on the room size and the taken measures to reduce  

 

hydrogen peroxide level in the treated room), Sanosil Q-Jet and room can be reused faster 

Disadvantages: Additional personal and time needed to air room, optionally an additional dehumidifier is 
 

needed, slight risk of new contamination through open windows 

 

 

 

Содержание Q-Jet Advanced 2 Series

Страница 1: ...Sanosil Q Jet Advanced 1 Operation and Maintenance Manual Sanosil Q Jet Advanced Series 2 Des M 56en Advanced V2 English...

Страница 2: ...Instructions 9 4 1 Warning 9 4 2 Room preparation positioning 9 4 3 Power Connection 10 4 4 Filling 10 4 5 Gun Nozzle 11 4 6 Switch on 11 5 Operation 12 5 1 Programs 13 5 2 Room Preparation 13 5 3 Ma...

Страница 3: ...ries Low risk Disregard can lead to medium or light injuries Disregard can damage thedevice Gives good advice for operating thedevice Further symbols will be explained thereafter 2 Safety DANGER Instr...

Страница 4: ...ble liquids Do always and only use Sanosil disinfection liquids 2 3 Danger and Caution Warnings The device must be used in full compliance with this Operation and Maintenance Manual It is strictly for...

Страница 5: ...n as needed basis depending on the volume of traffic and use Users should familiarize themselves with Q Jet operation by reading the Q Jet User Manual prior to first use It is essential that Q Jet use...

Страница 6: ...rical plug or parts are wet 8 Use only a Sanosil supplied electrical power cord 9 Take care to prevent water or other fluid from entering the device Should this occur allow to completely dry before us...

Страница 7: ...kg Weight filled 18 7 kg Noise level 84 dB 1m 3 2 Parts and Functions 1 Device 2 Power cord 3 ON OFF switch 4 Display 5 Air filter 6 Tank cap 7 Support for Gun Nozzle 8 Gun Nozzle 9 Service Flap 10 R...

Страница 8: ...place the device on top of trolley secure the device with the screw on the trolley CAUTION Risk of crushing Hands and fingers can be crushed between device and trolley Do always wear device on handle...

Страница 9: ...humidity the humidity must be lowered before start up or a lower dosage of the active substance must be selected Check temperature of the room to be between 15 to 35 C Affix warning board on entrance...

Страница 10: ...aplate Insert power cord 2 in IEC power socket 4 4 Filling Remove tank cap 6 Fill disinfection liquid max 4 2 liter Wipe up any spills with a clean paper towel or rag Do not tip the device on its side...

Страница 11: ...Advanced 11 Wind tank cap 6 until shut liquid level could be checked on the display 4 5 Gun Nozzle Remove cap from gun support 7 Insert gun 8 Gun must lock click 4 6 Switch on Tip ON OFF 3 to switch...

Страница 12: ...press i button Or button on remote control Display i button Return ENTER Number Arrow buttons buttons After switching on the device following is shown on thedisplay Tosee actual software version date...

Страница 13: ...om Preparation Close windows Pull back and secure all room dividing curtains If possible turn off then cover and seal all air handling units heaters and smokedetectors Check relative humidity of the r...

Страница 14: ...time Consumption in ml 10 3 min 99 ml 5 min 165 ml 6 min 198 ml 20 5 min 165 ml 8 min 264 ml 10 min 330 ml 30 7 min 231 ml 7 min 231 ml 14 min 462 ml 40 10 min 330 ml 11 min 363 ml 20 min 660 ml 50 1...

Страница 15: ...nseal HVAC vents and restart HVAC system to allow increasedairflow circulation in the room Once a safe ppm level is reached room can be re used again Shortening of waiting time after a minimal waiting...

Страница 16: ...filledsquares Toprogram days press ENTER For each day input starting time and confirm withENTER Toactivate day press I button square is filled when day is activated Device will start at the times pro...

Страница 17: ...proper personal equipment PPE including wraparound style goggles respirator mask filter class ABE1 filtering hydrogen peroxide for example 3M 4277 or 3M 7500 mask with 6057 filter and gloves should be...

Страница 18: ...Toget to the different cycles press every time ENTER Each cycle has an ON and an OFF time Please input for each cycle the desired time between 0 and 59 minutes and confirm with ENTER Enter 0 on cycle...

Страница 19: ...ase refer to the Sanosil test reports T319 T320 and T321 Room Size Fogging time consumption according the norm Afnor T72 281 2014 valid for Sanosil S015 for bacteria s yeasts molds 28ml m3 Fogging tim...

Страница 20: ...that the H2 O2 Level is below 1ppm Open windows and doors for min 15 minutes unseal HVAC vents and restart HVAC system to allow increasedairflow circulation in the room Once a safe ppm level is reache...

Страница 21: ...ime is displayed operating hours counter 5 9 Setup Language Chose mode SETUP LANGUAGE and confirm by pressing ENTER Tochoose language press respective number button 0 1 2 or 3 and confirm by pressing...

Страница 22: ...rrect on the display this calibration procedure should be done for initial use and frequently depending on the usage time by an authorized person 1 Place the device on a work bench 2 Do only use disti...

Страница 23: ...the filter will guarantee the proper functioning and lifetime of the Q Jet only Standard environment every 20 hours of operating Dusty risky environment every 10 hours of operating NOTE Any defects n...

Страница 24: ...on for the device is updated and available yearly Hydraulic Circuit Nozzle Visual status check cracks points of rupture bruising occlusions deposits daily Hydraulic Circuit Peristaltic pump tubing che...

Страница 25: ...ug device call service technician Device leaks liquid Possible internal damages Internal piping rupture or connections leaking Device was tipped on its side without completely empty Switch off and unp...

Страница 26: ...and fill it carefully Send complaint form with additional info and pictures if helpful to info sanosil com Sanosil Ltd forwards the complaint to the manufacturer The manufacturer decides whether the...

Страница 27: ...Sanosil Q Jet Advanced 27 10 CE Declaration of Conformity...

Страница 28: ...Sanosil Q Jet Advanced 28 Sanosil Ltd Eichtalstrasse 49 CH 8634 Hombrechtikon e mail info sanosil com www sanosil com Tel 41 55 254 00 54 Fax 41 55 254 00 59...

Отзывы: