8

OPTIONAL ACCESSORIES

Table 3 - Optional Accessories

Optional Accessories

Available For:

Guardian™ Ultraviolet Monitor - Analog/Digital

S17A through S50C

Sentry™ Safety Sensor

S17A through S50C

Steralert™

S17A through S50C

Audio Alarm

S17A through S50C

Elapsed Time Indicator Universal Input

S17A through S50C

Wall Mounting Kit

S17A through S50C

Time Delay Mechanism

S17A through S50C

Quantum Thermal Optimizer:

S17A through S50C

Solenoid Valve - Brass

S17A (3/4”)

S23A (3/4”)

S37C (1”)

S50C (1-1/2”)

 Sure Flo™ Flow Control Valve

S17A (3/4”)

S23A (3/4”)

S37C (1”)

S50C (1-1/2”)

Most optional accessories are
available for operation at 120v 60Hz
or 220v 50Hz. Please specify.
Consult Factory for 12v DC or other
special input power requirements.

 

Unless otherwise specifi ed PVC 
fl ow control valves are supplied. 
All PVC and Stainless Steel fl ow 
control valves are male NPT. 
Consult Factory for other fl ow 
control valves

MAINTENANCE

The water purifi er is designed to operate with a minimal amount of maintenance, providing the water quality does not exceed 
maximum concentration levels, see “Water Quality” in the “Product Application” section. Ordinary maintenance consists of;

• 

Lamp replacement is recommended every 10,000 hours of operation, approximately 12 months of continuous service.

• 

Cleaning of the quartz sleeve, when conditions warrant. It is recommended that the inspection of quartz sleeve be performed 
after one month of use. If quartz sleeve is found to be coated (not clear), then frequency of cleaning must be done more often. 
Deposits or discoloration on the surface of quartz sleeve are caused by excessive levels of the subject contaminant within 
the water that is in contact with the quartz sleeve. Most often false deposits on the quartz sleeve are caused by an excess of 
calcium (hardness), iron or manganese. Table 1 on Page 4 lists the maximum recommended concentration of these minerals 
in the water that passes through the ultraviolet purifi er. If you encounter diffi culty due to deposits on the quartz sleeve, your 
dealer will be able to recommend suitable pretreatment to reduce or eliminate the offending contaminant. If quartz sleeve is 
clean (clear) then frequency of cleaning may be extended.

 NOTE: 

SANITRON

® 

Ultraviolet water purifiers are equipped 

with a manual wiping mechanism making the process of routine cleaning easier and therefore, recommended weekly 
or at the very least monthly to insure your performance. NOTE: The use of optional Guardian™ Ultraviolet Monitor 
will detect loss of transmission due to coating on the quartz sleeve.

• 

Always disconnect the water supply and completely drain the water purifi er if it will be subjected to temperatures below 
freezing.

• 

Contact factory with questions.

I

NSPECTION

1.  Regularly inspect the water purifi er to ensure that the germicidal lamp is still in operation.

• On 

purifi ers not equipped with the Ultraviolet Monitor, lamp operation can be verifi ed by a visible glow through the 

translucent sight port. This provides an indication of lamp operation and does not indicate the level of ultraviolet 
intensity or transmission through the water.

• On 

purifi ers so equipped, the Guardian™ Ultraviolet Monitor provides visual indication of the ultraviolet intensity 

through the quartz sleeve and water in the purifi er chamber.

2.  To ensure proper operation of the water purifi er, regular biological testing should be performed on a schedule recommended 

by local public health authorities, or at minimum; at installation, quarterly for the fi rst year of service and annually, at lamp 
replacement, for the life of the water purifi er.

3.  Additional testing should be performed whenever modifi cations, change, or additions are made to plumbing system, pumps, 

well source water etc. to ensure adequate disinfection under new condition.

Содержание S17A

Страница 1: ...DXSSDXJH 1 D H PDLO LQIR XOWUDYLROHW FRP Extensive Product Information Available at ultraviolet com Document No 98 1377B9 August 2014 2014 Atlantic Ultraviolet Corporation Installation Operation Maintenance Read and Follow All Safety Instructions Save These Instructions Models S17A S23A S37C S50C Installation ULTRAVIOLET WATER PURIFIERS ...

Страница 2: ...ctions generally describe the installation operation and maintenance of the SANITRON line of water purifiers Models S17A S23A S37C and S50C Questions that are not specifically answerable by these instructions should be directed to the Factory Atlantic Ultraviolet Corporation takes all possible precautions when packaging equipment to prevent damage Carefully inspect and report all damages Do not in...

Страница 3: ... operate without proper electrical ground 12 Do not exceed water purifier s maximum rated flow capacity 13 Do not exceed maximum operating pressure of 100 PSI 14 Read and follow all notices and warnings on the water purifier 15 SAVE THESE INSTRUCTIONS This symbol pictorial is used to identify the need to wear protective gloves Caution indicates a POTENTIALLY hazardous situation which if not avoide...

Страница 4: ...is the product of ultraviolet intensity and the exposure time The operation of the SANITRON is as follows Water enters the purifier and flows into the annular space between the quartz sleeve and the chamber wall Suspended microorganisms are exposed to the ultraviolet rays emitted by the germicidal lamp The translucent sight port or optional ultraviolet monitor provides visual indication of germici...

Страница 5: ...d in an area where any possible condensation or leakage from the water purifier any purifier accessory and or plumbing will not result in damage to the area surrounding the water purifier For added protection it is recommended that a suitable drain pan be installed under the purifier The drain pan must be plumbed to an adequate free flowing drain to prevent water damage in event of a leak There ar...

Страница 6: ...upply which is required when removing the quartz sleeve 7 Unions should be installed on both the inlet and outlet of the water purifier this will allow easy removal of the water purifier from the plumbing if required Apply Teflon tape to threads of inlet and outlet ports to ensure a tight seal 8 When all plumbing connections are complete allow water to enter the water purifier at a low flow rate u...

Страница 7: ...tatus Alarm produces a high pitched pulsed tone when the water purifier is no longer functioning due to lamp or power failure 4 Sure Flo Flow Control Valve Limits water flow to the rated capacity of the purifier The flow control valve is located in line prior to the water purifier and should be protected from ultraviolet exposure by the use of a 90 degree elbow fitting between the flow control val...

Страница 8: ...on or manganese Table 1 on Page 4 lists the maximum recommended concentration of these minerals in the water that passes through the ultraviolet purifier If you encounter difficulty due to deposits on the quartz sleeve your dealer will be able to recommend suitable pretreatment to reduce or eliminate the offending contaminant If quartz sleeve is clean clear then frequency of cleaning may be extend...

Страница 9: ... wire away from the socket 3 Carefully withdraw lamp approximately 2 inches from chamber while feeding lamp socket and lead wire on opposite end of chamber 4 While holding lamp end carefully remove lamp socket on end now exposed 5 Next carefully slide lamp back into chamber until approximately 2 inches of the lamp is exposed on the opposite end Hold lamp and remove lamp socket 6 Lamp should now be...

Страница 10: ... is easily accomplished using the dual action wiper mechanism 1 Lift wiper clip up and away from wiper rod 2 Grasp wiper knob and gently pull out away from the purifier until it reaches its stop 3 Push wiper back in toward the purifier until it reaches its second stop 4 Repeat steps 2 and 3 as necessary 5 Holding wiper in place return wiper clip and snap over wiper rod in front of the wiper lock s...

Страница 11: ... washer from static gland nut before proceeding 6 Carefully remove quartz sleeve from chamber NOTE It is advisable to support the quartz sleeve on the opposite end with your finger so that it does not drop to the bottom of the chamber as it slides into the chamber 7 Once the quartz sleeve is removed clean with alcohol or a mild non abrasive detergent Stubborn stains usually can be removed with a d...

Страница 12: ...16 wrench to loosen and remove the 1 4 nut from the head clamp 4 Withdraw chamber head and wiper assembly from the chamber 5 Any broken pieces of the quartz sleeve can now be removed through the open end of the purifier chamber Flush water through chamber being careful to remove all quartz fragments from the interior of the chamber 6 Carefully discard all pieces of the broken quartz sleeve 7 Inspe...

Страница 13: ...nspected and can be replaced if worn or damaged 8 Reassemble replacing the o rings quartz rod and securing in place with the quartz rod sensor probe cap Tighten the quartz rod sensor probe cap by hand only DO NOT USE HAND TOOLS Tightening with hand tools may damage the quartz rod or o ring seal 9 Reinstall sensor probe into the center fitting of the chamber and hand tighten 10 Slowly restore water...

Страница 14: ...e connections or fittings Tighten suspect connection or fitting or shut down purifier drain and remove fitting or connection Clean threads reapply thread sealing tape and reinstall Poor purifier performance AND OR Low UV intensity As indicated on optional GuardianTM Ultraviolet Monitor Quartz sleeve fouled Clean quartz sleeve see Quartz Sleeve Cleaning in the Maintenance Section Sensor Probe if eq...

Страница 15: ...in Plug 1 4 NPT 1 4 NPT 1 4 NPT 1 4 NPT Lamp Out Indicator Translucent Sight Port Translucent Sight Port Translucent Sight Port Translucent Sight Port Ultraviolet Monitor Optional Optional Optional Optional Flow Control Valve Optional Optional Optional Optional Audio Alarm Optional Optional Optional Optional Solenoid Valve Optional Optional Optional Optional Time Delay Mechanism Optional Optional ...

Страница 16: ...16 REPLACEMENT PARTS SANITRON MODELS S17A S23A Figure 9 Exploded View S17A S23A ...

Страница 17: ...1524D1 1 25 1523E1 16 Wiper Knob 1 25 1222 1 25 1222 17 Wiper Lock Spacer 1 25 1512A1 1 25 1512A1 18 Wiper Clip 1 25 1507C1 1 25 1507C1 19 Dynamic Gland Wiper Lock 1 25 1510C1 1 25 1510C1 20 Drain Plug 1 27 1216 1 27 1216 21 Screw No 8 x 3 8 long 4 50 1323 4 50 1323 22 Speed Nut 6 32 2 50 1314 2 50 1314 23A Ballast 120v 60Hz 1 10 0137 1 10 0137 23B Ballast 220v 50Hz 1 10 0136 1 10 0136 24 Screw Mo...

Страница 18: ...18 SANITRON MODELS S37C S50C Figure 10 Exploded View S37C S50C ...

Страница 19: ...25 1512A1 18 Wiper Clip 1 25 1507C1 1 25 1507C1 19 Dynamic Gland Wiper Lock 1 25 1510C1 1 25 1510C1 20 Drain Plug 1 27 1216 1 27 1216 21 Screw 1 4 20 x 3 8 long 4 50 1034 4 50 1034 22 Lock Washer 1 4 4 50 1321 4 50 1321 23 Flat Washer 1 4 4 50 1317 4 50 1317 24 Ballast Housing Cover 1 25 0394A 1 25 0350A 25 Screw No 8 x 3 8 long 8 50 1323 8 50 1323 26 Speed Nut 6 32 2 50 1314 2 50 1314 27A Ballast...

Страница 20: ... has been made to determine the patent status of applications illustrated or described in this publication Inclusion in this publication of any design or method of use which may be patented is not to be construed as promoting or sanctioning unauthorized use Atlantic Ultraviolet Corporation makes every effort to ensure that the SANITRON Ultraviolet Water Purifiers are products of superior quality a...

Отзывы: