background image

DT 300 

(v0.1)

 

IS DT300 1410 (SE_GB_D_ES).doc 

 

 

GB 

 

 

 

Altering of set point, and adjustments shall be carried out 

by trained personnel only, who has adequate knowledge of the 
products. Incorrect handling may cause that the system 
becomes inoperative.

 

 

 

DT300 is an instrument to be used with: 

- Detectors type G-series 

- Detectors type MP-series in combination with SPU/MPU 

- Monitoring units type MPU & SPU 

 

DT300 Instrument can be used as follows: 

-Determine the environment of the actual detector. 
Units with SC-sensors only.

 

- Verify offset-value when changing sensor head 
Units with SC-sensors only. 
- Measure sensor signal during ”bump-test”, or when using 
reference gas 
- Check of alarm 

thresholds

 

 

1. Start of DT300 

The instrument is switched on via the on/off button on the left 
side of the unit. At start up the letter “W” will be flashing on the 
left side of “Ref sensor” value. The letter “W” indicates that the 
sensor is warming up. Check that this “Ref-value” is stabilized 
before to use the instrument. No adjustment is to be carried out 
before that the indication has 

extinct.

 

 

 

 

 

 

2. Control of the Ref-sensor offset (SM-300). 

To control the status of the Ref-sensor, one should start the 
instrument in a possible clean environment, e.g. outside.  

The “zero”-value of the sensor is defined as the signal 

(VDC) when in clean air

  

By that, it is normal that the signal varies in various 
environments, and also due to temperature and humidity.

 

The output is normally higher during summer than winter. 
The acceptable variation is given in the data sheet of the 
specific sensor. The factory set sensor module, does normally 
not require any adjustment. Although, a small “drift” can occur 
due to environment and the frequency of use. 

If the value shown in the display is within the green area

 , 

no adjustments is required

 

If the value shown in the display is within the orange area

the value should be adjusted

 

If the value shown in the display is within the red area, the 
sensor has to be replaced 

 

If the value in the display increases when entering a 

plant room e.g., this indicates that there are substances 

present that effects the sensor. This value is to be the 

reference value for the detectors within this perimeter. 

 

 

 
 

Normal variation 

 

Adjustable deviation 

 

Sensor to be replaced 

 

3. Adjustment of the reference sensor offset (SM300) 

Always start the procedure by checking the due date of the SM-
module. (See label on the module) 

 If due date is passed, the SM-module shall be replaced 

without delay. Sensors is a perishable item that is effected by 
dust, dirt and can be “poisoned” by various substances such as, 
aerosols, paint, solvent fumes, glue and similar. 
- Before adjustment, the unit should be active in a stable 
environment for at least one hour

 

At the SM-module there is a potentiometer that is accessible by 
the hole in the end of the housing of the instrument.  
(See picture below) 
Carefully adjust by using a small screwdriver. When the offset 
value passes from the orange area to the green area, called 
(normal variation) of the sensor, the task is completed. 
The instrument is now ready for use.  

 

 

 

4. Connect the instrument to the test-socket of the detector 

The small guide at the cable-socket should be pointing to the 
right, when the three potentiometers for alarm threshold 
adjustment of the actual unit, are located underneath the test 
terminal socket. (See picture, page 2) 

- When testing a detector of the MP-series without the six 

pin test socket, an adaptor cable is required to enable 

connection to the DT300. 

 

 

 

5. Check the system voltage of the detector 

The system voltage of the detector shall be 5 VDC ± 0,15V 
(See picture, page 1) 
If the deviation is larger, check the power supply of the detector. 

 

6. Check the detector output 

 

Make sure that the sensor head of the SM-module and the 

detector are identical. 

- The detector has to be powered for at least one hour 

before control is carried out, and it should not be exposed 

to draft.  

The sensor output of the connected detector is visualised at the 
very left of the upper row of digits. This value is to be compared 
with the value of the reference sensor within the DT300. Adjust 
the value of the detector via the potentiometer, marked “GV-
offset pot”

 

(see picture, page 2) to reach the similar value as of 

the ref-sensor.  See the instruction of the specific detector. 

 

 At large deviations, > +-0,5VDC, the sensor head is 

probably damaged/poisoned, and shall be replaced. 

 

 

Guide 

SM-module adjustment 

Содержание DT 300

Страница 1: ...0 SE_GB_D_ES doc 1 SE Instruktioner GB Instructions D Betriebsanleitung ES Instrucciones DT 300 SE GB Diagnostic Tool Detector signal offset System voltage Reference sensor signal offset Alarm C Batte...

Страница 2: ...00 v0 1 IS DT300 1410 SE_GB_D_ES doc 2 Detektorer Detectors Detektoren Detectores G serie 24V 230V Centraler Monitoring units Kontrolleinheiten Central MPU2C 4C 6C SPU24 230 Terminal guide Terminal gu...

Страница 3: ...r f r Ammoniak NH3 1000 60 136 SM300 NH3 4000 Sensor f r Ammoniak NH3 4000 60 137 SM300 NH3 10000 Sensor f r Ammoniak NH3 10000 GB The instrument consists of two parts a base unit with display batteri...

Страница 4: ...rs v rde kar n r man g r in i tex ett maskinrum visar detta att i lokalen finns gaser mnen som p verkar sensorn Visat v rde blir d referens f r de detektorer som sitter i den milj n Normal variation J...

Страница 5: ...rva 8 Kontroll av sensorn respons och terh mtningstid genom s k bumptest Efter kontroll och eventuell justering av sensor offset skall en funktionskontroll g ras Genom att f ra p en liten m ngd koncen...

Страница 6: ...or This value is to be the reference value for the detectors within this perimeter Normal variation Adjustable deviation Sensor to be replaced 3 Adjustment of the reference sensor offset SM300 Always...

Страница 7: ...and eventual adjustments of the sensor off set a function control is to be carried out By exposing the sensor to a small amount of concentrated gas directly to the sensor head the response and recover...

Страница 8: ...eit auftreten kann Ist der Wert auf dem Display in dem gr nen Bereich sind keine Einstellungen n tig Ist der Wert auf dem Display im orangenem Bereich sollte der Wert angepasst werden Ist der Wert auf...

Страница 9: ...esen werden Um den entsprechenden PPM Wert f r einen Schwellenwert in VDC zu finden m ssen Sie zuerst den VDC Wert auf der senkrechten Achse finden Anschlie end folgen Sie in einer wagerechten Linie b...

Страница 10: ...chte anfangen zu blinken muss die Batterie ausgetauscht werden 10 St rungsbeseitigung Sollte der aktuelle Detektor keinen Alarm ausl sen wenn er Gas ausgesetzt ist berpr fen Sie bitte Folgendes GV Sig...

Страница 11: ...de la zona verde no se requieren ajustes Si el valor mostrado en el display est dentro de la zona naranja el valor debe de ser ajustado Si el valor mostrado en el display est dentro de la zona roja e...

Страница 12: ...asta que alcanza el gr fico y all despu s de seguir una l nea vertical imaginaria hasta que alcanza el eje horizontal del diagrama En el eje horizontal se puede leer el valor PPM correspondiente Si se...

Страница 13: ...a gas comprobar lo siguiente GV offset y se al del sensor en base a los puntos 6 y 8 Configuraci n de los umbrales de alarma Configuraci n de los DIP switches para el retardo de la alarma Ese detecto...

Страница 14: ...DT 300 v0 1 IS DT300 1410 SE_GB_D_ES doc 14...

Отзывы: