SAM-III Construction Manual
See last page for document and copyright information, File: SAM-III Magnetometer Construction.doc, Page 70
Temperature sensor test:
After installing the digital thermometer, turn on power to the SAM-III and then run SAM_VIEW. Go to
the directory or folder where SAM_VIEW stores the data (specified in Edit – Setup – Logging – Data
Directory). After a moment, some .txt and .log files will appear. Open the sam_data.txt file and look in
the last column to the right of the Z-sensor data values. The temperature should be shown. See example
below. Gaps in the temperature data are normal.
DD.MM.YY HH:MM:SS: X,DDDD,Y,DDDDD,Z,DDDDD,TT.T
28.07.10 22:13:58: X,6809,Y,44871,Z,-7992,22.8
28.07.10 22:14:00: X,6809,Y,44871,Z,-7992,22.8
28.07.10 22:14:02: X,6809,Y,44871,Z,-7992,22.8
28.07.10 22:14:04: X,6809,Y,44871,Z,-7992
28.07.10 22:14:06: X,6809,Y,44871,Z,-7992,22.8
28.07.10 22:14:08: X,6808,Y,44871,Z,-7992,22.8
28.07.10 22:14:10: X,6808,Y,44871,Z,-7992,22.8
28.07.10 22:14:13: X,6808,Y,44868,Z,-7992
28.07.10 22:14:15: X,6808,Y,44868,Z,-7992,22.7
28.07.10 22:14:17: X,6808,Y,44868,Z,-7992
28.07.10 22:14:19: X,6808,Y,44868,Z,-7992,22.7