14
8
WORKING ENVIRONMENT AND STORAGE
The device cannot be used in environments that aren’t conform to the specifications given in paragraph
ENVIRONMENTAL CHARACTERISTICS (3.2)
In accordance with CEI EN 60601-1-11 sections 1.1 and 3.2 can be used:
•
PROFESSIONAL HEALTHCARE STRUCTURES (hospitals, medical offices, surgical
centers, dental clinics, birth centers, health care facilities, health care facilities
for multiple treatment and emergency, medical services);
•
HOME HEALTH ASSISTANCE (home / place in which a patient lives or other
places where there are patients).
WARNINGS
PREVIOUS POINT LIMITATION: THE DEVICE CANNOT BE USED IN MOTION (EN 60601-1-11 Point 3.6)
8.1
Tables guide of ELECTROMAGNETIC COMPATIBILITY
In accordance with CEI EN 60601-1-2, section 6.8.3.201 (letter "a" and "b") provides information on various aspects of
electromagnetic compatibility and in particular the methods of carrying out the type tests carried out and the characteristics
required for the environment in wichthe device can be used.
Electromagnetic Emissions
The device is intended for use in the electromagnetic environment specified below.
WARNINGS
The user must ensure that it is used in that specific environment.
Failure to observe may generate risk situations.
Emissions Test
Conformity
Electromagnetic Environment - Guidance
CEI EN 55011
Emissions RF
compliant
(group 1)
The device uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
CEI EN 55011
Emissions RF
compliant
(class B)
The device is suitable for use in all rooms including domestic
establishments and those directly connected to a power grid at
low voltage which supplies buildings used for domestic
purposes.
CEI EN 61000-3-2
Harmonic Emissions
compliant
(class A)
CEI EN 61000-3-3
Emissions ofvoltage
fluctuations/flicker
compliant
This first table (rif.201 of the standard cited above) realizes the aspects related to the emissions of the device, and all the possible
noise or interference generated by the device that could disturb other electronic equipment more or less close.